Ohio Medical care-e-vac 3 Operator And  Maintenance Manual

Ohio Medical care-e-vac 3 Operator And Maintenance Manual

Portable suction aspirator

Advertisement

PORTABLE SUCTION ASPIRATOR
PART NUMBER 758000
OPERATOR / MAINTENANCE
M
A
N
U
A
L

Advertisement

Table of Contents
loading

Summary of Contents for Ohio Medical care-e-vac 3

  • Page 1 PORTABLE SUCTION ASPIRATOR PART NUMBER 758000 OPERATOR / MAINTENANCE...
  • Page 2: Table Of Contents

    The medical device may be powered by a 120 VAC outlet or 12 VDC source. It is equipped with an internal battery. The primary intended use of the care-e-vac 3 is to aspirate saliva, mucous, vomitous or other aspirant from the mouth and or airway to allow adequate respiration or ventilation of the patient.
  • Page 3: Definitions And Symbols Legend

    Amber Light = CHARGING / AC. Unit is properly connected to the power source. This light will remain illuminated as long as the unit is connected to the power source. Ohio Medical, 1111 Lakeside Drive, Gurnee, Il 60031 (866) 549-6446 | Fax: (847) 855-6218 www.ohiomedical.com...
  • Page 4: Instructions For Use

    Set the vacuum level to the least amount of vacuum necessary to properly suction the patient. Ohio Medical, 1111 Lakeside Drive, Gurnee, Il 60031 (866) 549-6446 | Fax: (847) 855-6218 www.ohiomedical.com...
  • Page 5: Maintenance

    WARNING: Remove device from A/C power source prior to cleaning. Ohio Medical, 1111 Lakeside Drive, Gurnee, Il 60031 (866) 549-6446 | Fax: (847) 855-6218 www.ohiomedical.com —– 550542 (Rev.6) 06/2016...
  • Page 6 Ohio Medical recommends changing the battery every 3 years, depending on the frequency of use. Ohio Medical, 1111 Lakeside Drive, Gurnee, Il 60031 (866) 549-6446 | Fax: (847) 855-6218 www.ohiomedical.com —– 550542 (Rev.6) 06/2016...
  • Page 7 Items that can be cleaned: Housing, Gauge and Regulator b. Items that must be replaced if exposed to fluids and contaminates: Tubing, Control Circuit, Power Supply, Pump, Switch. Ohio Medical, 1111 Lakeside Drive, Gurnee, Il 60031 (866) 549-6446 | Fax: (847) 855-6218 www.ohiomedical.com —– 550542 (Rev.6) 06/2016...
  • Page 8: Troubleshooting / Removal Of Service Components

    2. Blockage in vacuum lines. 2. Ensure the vacuum regulator is not com- pletely turned off and that there are no obstructions in the line. Ohio Medical, 1111 Lakeside Drive, Gurnee, Il 60031 (866) 549-6446 | Fax: (847) 855-6218 www.ohiomedical.com —– 550542 (Rev.6) 06/2016...
  • Page 9: Device Specifications

    6 hours or less to 85% charging, depending on initial charge 8 hours full charge time Charge Method: 110/120 VAC outlet Ohio Medical, 1111 Lakeside Drive, Gurnee, Il 60031 (866) 549-6446 | Fax: (847) 855-6218 www.ohiomedical.com —– 550542 (Rev.6) 06/2016...
  • Page 10: 800 Cc Canister Instructions

    The canister is NOT intended to be used as a measuring device. It should ONLY be used for general reference and should NEVER be used for specific measurement. • Canister contents are considered potentially hazardous. Use appropriate Personal Protective Equipment (PPE) and handle accordingly. Ohio Medical, 1111 Lakeside Drive, Gurnee, Il 60031 (866) 549-6446 | Fax: (847) 855-6218 www.ohiomedical.com —– 550542 (Rev.6) 06/2016...
  • Page 11: Osha Hazard Communications Standard

    If there are any questions pertaining to this subject, please call our Customer Service Department at (866) 549-6446. Thank you for your cooperation. OHIO MEDICAL 1111 Lakeside Drive Gurnee, Illinois 60031 Ohio Medical, 1111 Lakeside Drive, Gurnee, Il 60031 (866) 549-6446 | Fax: (847) 855-6218 www.ohiomedical.com —– 550542 (Rev.6) 06/2016...
  • Page 12: Warranty

    Standard surface freight shipping costs to return the unit to the original purchaser will be paid by OHIO MEDICAL. Any product returned must be free from contamination by toxic or hazardous substances in accordance with OSHA HAZARD COMMUNICATION STANDARD 29 CFR, Section 1910.1200 and OSHA BLOOD BORNE...
  • Page 13: Accessories

    CEV3 Pump Kit GAST WARNING: The use of accessories, service parts, and/or cables, other than those listed above, may result in Ohio Medical, 1111 Lakeside Drive, Gurnee, Il 60031 (866) 549-6446 | Fax: (847) 855-6218 www.ohiomedical.com —– 550542 (Rev.6) 06/2016...
  • Page 14: Emc Information

    A used battery is a valuable resource. Insulate the battery terminals and dispose the battery by taking it to a battery recycling center. Ohio Medical, 1111 Lakeside Drive, Gurnee, Il 60031 (866) 549-6446 | Fax: (847) 855-6218 www.ohiomedical.com —– 550542 (Rev.6) 06/2016...
  • Page 15: Electromagnetic Compatibility Declarations

    XV. Electromagnetic Compatibility Declarations for care-e-vac 3 Guidance and Manufacturer's Declaration - Electromagnetic Emission's The care-e-vac 3 is intended for use in the electromagnetic environment specified below. The user of the care-e-vac 3 should assure that it is used in such an environment. Emissions Test...
  • Page 16 Ohio Medical (866) 549-6446 | Fax: (847) 855-6218 1111 Lakeside Drive, Gurnee, IL 60031 www.ohiomedical.com 550542 (Rev.6) 06/2016...