Geokon 8020-59 Instruction Manual page 25

Vibrating wire to analog converter
Table of Contents

Advertisement

19
Typing "S" <Enter> displays the 8020-59 status:
*S
Monitor Enabled
Configuration 2:
16 VW's and 16 Thermistors
1,5000,1000,+,Enabled
2,25000,160,+,Disabled
3,25000,160,+,Disabled
4,25000,160,+,Disabled
5,25000,160,+,Disabled
6,25000,160,+,Disabled
7,25000,160,+,Disabled
8,25000,160,+,Disabled
9,25000,160,+,Disabled
10,25000,160,+,Disabled
11,25000,160,+,Disabled
12,25000,160,+,Disabled
13,25000,160,+,Disabled
14,25000,160,+,Disabled
15,25000,160,+,Disabled
16,25000,160,+,Disabled
Typing "R" <Enter> will return the firmware revision.
Typing "Z" <Enter> will put the 8020-59 into low power sleep mode (off), drawing less than
10μA from the 12V power source (less than 16.5mA from the 24V power source).
Finally, pressing the <Esc> key (in C1 single channel mode only) will start the readings screen
with the 8020-59 outputting 0-5V and 4-20mA:
1,11665.75,22.8,-,1,#
1,11665.95,22.7,-,1,#
1,11666.50,22.8,-,1,#
1,11666.60,22.8,-,1,#
1,11667.05,22.8,-,1,#
1,11666.55,22.9,-,1,#
1,11666.10,22.8,-,1,#
1,11665.65,22.8,-,1,#
1,11666.30,22.9,-,1,#
1,11666.50,22.8,-,1,#
1,11666.65,22.8,-,1,#
1,11666.35,22.8,-,1,#
1,11665.85,22.8,-,1,#
1,11666.10,22.8,-,1,#
1,11666.60,22.8,-,1,#
Where;
Comma delimited field one is the channel number, followed by the reading in digits, then the
transducer temperature in deg. C, the slope, and finally the gage type. The pound sign (#)
signifies the end of the data string.

Advertisement