Example:
(Assumes the use of default values and transducer type of 4500S-100):
Minimum Limit = 0 digits
Maximum Limit = 25,000 digits
Digits Span = 25,000 – 0 = 25,000 digits
Vout = 3.250V
Digits = Maximum Limit – (Vout * (Digits Span/5V))
Digits = 25,000 – (3.25V * (5000 digits/V))
Digits = 25,000 – 16,250
Digits = 8750

4.4 Connection to a Voltage Input DAS

The 8020-59 outputs a voltage (V OUT, 0-5V) that is proportional to the digits of the transducer
being read (ENABLE = 0V (or disconnected) in Single Channel Mode, after the first of two
CLOCKS in 16-Channel Mode, or after each CLOCK in 32-Channel Mode). The 8020-59 offers
a digits to voltage output with 20-bit resolution (one part in 1,048,576) of the span of the
transducer (default span is 25,000 digits). This output voltage can be input directly to any Data
Acquisition System capable of reading a 0-5V analog voltage.
The 8020-59 also outputs a voltage proportional to the temperature (also on V OUT) of the
transducer being read (ENABLE = 5V in Single Channel Mode or after the second of two
CLOCKS in 16-Channel Mode). The 8020-59 offers a temperature voltage output with 10-bit
resolution (one part in 1024) over a 100º C span (-20º C to +80º C). See Section 4.6 for specific
calculations.

4.5 Connection to a Current Input DAS

The 8020-59 sources a current (I-OUT, 4-20mA) that is proportional to the digits of the
transducer being read (ENABLE = 0V (or disconnected) in Single Channel Mode, after the first
of two CLOCKS in 16-Channel Mode, or after each CLOCK in 32-Channel Mode). The 8020-59
offers a digits current output with 20-bit resolution (one part in 1,048,576) of the span of the
transducer (default span is 25,000 digits). This output current can be input directly to any Data
Acquisition System capable of reading a non-isolated 4-20mA analog current.
The 8020-59 also outputs a current proportional to the temperature (also on I-OUT) of the
transducer being read (ENABLE = 5V in Single Channel Mode or after the second of two
CLOCKS in 16-Channel Mode). The 8020-59 offers a temperature current output with 10-bit
resolution (one part in 1024) over a 100º C span (-20º C to +80º C). See Section 4.6 for specific
calculations.
See Appendix C for the various voltage and current measurement options.
15