Page 1
Vibrating Wire to Analog Converter No part of this instruction manual may be reproduced, by any means, without the written consent of Geokon, Inc. The information contained herein is believed to be accurate and reliable. However, Geokon, Inc. assumes no responsibility for errors, omissions, or misinterpretation.
Page 3
Upon examination by Geokon, if the unit is found to be defective, it will be repaired or replaced at no charge. However, the WARRANTY is VOID if the unit shows evidence of having been tampered with...
TABLE of CONTENTS 1. INTRODUCTION ..............................1 2. INSTALLING AND CONFIGURING THE 8020-59 SOFTWARE APPLICATION ........2 2.1 C COM P ........................... 5 ONFIGURING THE 3. QUICK START (SINGLE CHANNEL) ....................... 7 4. SINGLE AND MULTI-CHANNEL OPERATION ....................10 4.1 T .............................10...
Page 5
IGURE NSTALL IZARD NSTALLATION OMPLETE 8 - I 8020-59 A ..................5 IGURE NSTALL IZARD AUNCH PPLICATION 9 - 8020-59 S .................. 5 IGURE OFTWARE PPLICATION TARTUP IALOG 10 - C ....................6 IGURE OMMUNICATIONS ARAMETERS IALOG 11 - S ....................
Page 6
TABLES 1 - C ..................1 ABLE ONNECTOR INOUT AND IGNAL ESCRIPTION 2 - T ..................28 ABLE HERMISTOR ESISTANCE ERSUS EMPERATURE 3 - E ..................42 ABLE NGINEERING NITS ULTIPLICATION ACTORS EQUATIONS 1 - V ................14 QUATION IBRATING REQUENCY TO IGITS ONVERSION 2 - A ........................14...
1. INTRODUCTION The Model 8020-59 Vibrating Wire (VW) to Analog Converter is a low cost module that provides a simple way to connect Geokon’s vibrating wire transducers to data acquisition systems that are not capable of reading frequency signals nor able to generate the proper signals required to excite VW transducers.
APPLICATION A free setup and configuration application is available for download on Geokon’s website, providing a user-friendly way to set up the 8020-59 VW to Analog Converter. Follow the steps below to install the software: 1) After downloading the installer from the Geokon website, right-click on the zip file, “8020- 59_setup.zip”, and select “Extract All…”...
Figure 3 - Extracted 8020-59 Installer 4) After a few moments, the dialog shown in Figure 4 will be displayed. Figure 4 - Install Wizard, Start Screen 5) After clicking on the “Next >” button, the dialog shown in Figure 5 is displayed.
6) Choose whether to install the 8020-59 Software Application for all users or just for the current user (see Figure 5). NOTE: Without administrative privileges on the PC, the only option allowed will be “Install just for me”. After making a selection, click “Next >”; this will bring up the dialog shown in Figure 6.
After launching the 8020-59 Software Application, the first thing that must be done is to configure the COM Port. Click on the “Com Setup” (or press the “Alt” and the “c” key at the same time) to select a COM port for communication with the 8020-59 VW to Analog Converter (See Figure 9).
(such as COM1) or an external one via a USB to Serial Converter (see Figure 10). Some newer PCs are so fast that a small delay between characters is needed to give the 8020-59 time to process the characters. The “Inter-character Delay” is specified in milliseconds.
VW Transducer VW Transducer Calibration Report 1) Install the 8020-59 software on the PC that will be used to interface to the 8020-59 unit (see Section 2). 2) Connect the 8020-59 to the computer’s serial port (typically COM1). If using an external USB to Serial converter, drivers for this device will need to be installed before this step (see Section 2.1).
“Do you want to upload the current settings to the 8020-59?”. Click the “Yes” button. The Single Channel Monitor screen should now appear (see Figure 12), and after several seconds should update with a reading.
Page 15
See Appendix D for further information on Data Reduction. 12) Click “Exit” (or press the “Alt” and “e” keys) to return to the Setup Screen. After setup is complete, the PC is disconnected from the 8020-59 and the sensor is in a pressurized state, the “Current Reading” (R ) “Digits”...
Acquisition System to select either 1 of 16 or 1 of 32 transducers. The 8020-59 was developed with multiplexing in mind and uses the same control signals as the Geokon model 8032 Vibrating Wire Multiplexer (for control signal timing requirements, refer to the model 8032 Multiplexer Instruction Manual).
4.2 Operating Modes The 8020-59 has three modes of operation: single channel mode, 16-Channel mode, and 32- Channel mode. Each operation mode is detailed in the subsections below. 4.2.1 Single Channel Mode In this mode (Figure 13), the 8020-59 will maintain continuous excitation of the VW transducer and provide a continuous 0-5V and 4-20mA output to the host DAS, updated approximately every second.
When the final transducer has been read, the host DAS brings CLOCK and ENABLE both low (0V) to reset the system. The 8020-59 will go back to sleep until it is time for the next reading. Using this feature, up to 16 Vibrating Wire transducers with thermistors may be multiplexed into a single DAS.
When the final transducer has been read, the host DAS brings CLOCK and ENABLE both low (0V) to reset the system. The 8020-59 will go back to sleep until it is time for the next reading. Using this feature, up to 32 Vibrating Wire transducers may be multiplexed into a single DAS.
Digits = [Frequency (Hz)]² x 0.001 Equation 1 - Vibrating Wire Frequency to Digits Conversion The 8020-59 provides analog outputs that are directly proportional to digits. Referring to the sample calibration in Figure 36 of Appendix E, a value in digits can be derived from the voltage and current outputs as described below.
4.4 Connection to a Voltage Input DAS The 8020-59 outputs a voltage (V OUT, 0-5V) that is proportional to the digits of the transducer being read (ENABLE = 0V (or disconnected) in Single Channel Mode, after the first of two CLOCKS in 16-Channel Mode, or after each CLOCK in 32-Channel Mode).
4.6 Temperature Measurement The 8020-59 is capable of reading the thermistor that is part of a VW sensor. The 8020-59 outputs this reading as a voltage (0-5V) and current (4-20mA) that are proportional to temperature whenever ENABLE is high (5V), in Single Channel Mode, or while the second clock pulse is high, in 16-Channel mode.
5. COMMUNICATIONS The 8020-59 offers a standard 9-pin RS-232 Serial Port for connection to most desktop and laptop computers (an optional USB to Serial interface adapter is available). If using a terminal emulator program such HyperTerminal, Putty, etc., for command line set up (see Section 5.1) of the 8020-59, configure the program’s communication parameters as follows:...
Page 24
<Esc> Exit and Start Taking Readings (C1 only) Commands C1, C2 and C3 configure the 8020-59 for single channel, 16-Channel multiplexing or 32-Channel multiplexing, respectively. Note: The default value for the lower limit is zero, so for cases where this is acceptable, there is no need to explicitly set the lower limit.
Page 25
16,25000,160,+,Disabled Typing “R” <Enter> will return the firmware revision. Typing “Z” <Enter> will put the 8020-59 into low power sleep mode (off), drawing less than 10μA from the 12V power source (less than 16.5mA from the 24V power source). Finally, pressing the <Esc> key (in C1 single channel mode only) will start the readings screen with the 8020-59 outputting 0-5V and 4-20mA: 1,11665.75,22.8,-,1,#...
Startup dialog (see Figure 9). From the Startup dialog, the major functions (Mode, Communications, and Calibration) of the 8020-59 can be accessed. Click Com Setup to select the Com Port that the 8020-59 is connected to (see Section 2.1 and Figure 10).
6.4 Single Channel Configuration If the 8020-59 is to be used with only one transducer, the 8020-59 Mode should be set to Single Channel. Click “Single Channel” (or press the “Alt” and “s” keys) to display the Single Channel Configuration window (see Figure 11).
6.5 16-Channel Configuration The 8020-59 can be used in conjunction with the Geokon model 8032 Multiplexer to configure and monitor up to 16 vibrating wire transducers with thermistors. For wiring details of this configuration, see Appendix C. The 8020-59 Mode should be set to 16-Channel. Clicking “16- Channel”...
Page 29
In the Transducer Limits section, the current minimum and maximum settings for the channels stored in the 8020-59’s on-board EEPROM are in the gray fields on the right. New values in the configuration are in the white fields on the left. Enter the transducer’s limits (in digits) from the transducer’s calibration report at “Maximum (digits)”...
The screen updates based on ENABLE and CLOCK timing from the Data Acquisition System. If monitoring the 8020-59 outputs is not of interest, then click Activate/Exit from the Configuration Window to put the 8020-59 into low power Standby (Sleep) mode, waiting...
6.6 32-Channel Configuration 32-Channel Configuration and Monitor modes work identically to their 16-Channel counterparts with the only difference being that there are no temperature measurements (see Figure 20). Figure 20 - 32-Channel Monitor Screen Figure 21 - Channel Three Monitor...
APPENDIX A. SPECIFICATIONS A.1 8020-59 Specifications Physical: Dimensions (L x W x H): 111.13 mm, 108.36 mm x 36.53 mm 4.375" x 4.266" x 1.438" Weight: 0.456 lbs. Operating Temperature: -20 to +80 degrees Celsius Power Requirements: +12V Input Voltage Range:...
APPENDIX B. THERMISTOR TEMPERATURE DERIVATION Thermistor Type: YSI 44005, Dale #1C3001-B3, Alpha #13A3001-B3 Resistance to Temperature Equation: A+B ( LnR ) +C(LnR) -273.2 Equation 8 - Resistance to Temperature Where; T = Temperature in °C. LnR = Natural Log of Thermistor Resistance A = 1.4051 ×...
APPENDIX C. WIRING CONFIGURATIONS C.1 Voltage Output – Single Channel Mode (Digital I/O DAS) Figure 24 - Connection Example for One Channel 8020-59 with Voltage Output to Digital I/O DAS...
C.2 Voltage Output – 16-Channel Mode (Digital I/O DAS) Figure 25 - Connection Example for 16-Channel Multiplexer, 8020-59 with Voltage Output and Digital I/O DAS...
C.3 Voltage Output – 32-Channel Mode (Digital I/O DAS) Figure 26 - Connection Example for 32-Channel Multiplexer, 8020-59 with Voltage Output and Digital I/O DAS...
C.4 Current Output – Single Channel Mode (Digital I/O DAS) Figure 27 - Connection Example for One Channel 8020-59 with Current Output to Digital I/O DAS...
C.5 Current Output – 16-Channel Mode (Digital I/O DAS): Figure 28 - Connection Example for 16-Channel Multiplexer, 8020-59 with Current Output and Digital I/O DAS...
C.6 Current Output – 32-Channel Mode (Digital DAS) Figure 29 - Connection Example for 32-Channel Multiplexer, 8020-59 with Current Output and Digital I/O DAS...
C.7 Voltage Output – Single Channel Mode (PLC DAS) Figure 30 - Connection Example for One Channel. 8020-59 with Voltage Output, 8020-59 PLC and PLC DAS...
C.8 Voltage Output – 16-Channel Mode (PLC DAS) Figure 31 - Connection Example for 16-Channel Multiplexer, 8020-59 w/ Voltage Output, 8020-59 PLC and PLC DAS...
C.9 Voltage Output – 32-Channel Mode (PLC DAS) Figure 32 - Connection Example for 32-Channel Multiplexer, 8020-59 w/ Voltage Output, 8020-59 PLC and PLC DAS...
C.10 Current Output – Single Channel Mode (PLC DAS) Figure 33 - Connection Example for One Channel 8020-59 with Current Output, 8020-59 PLC and PLC DAS...
APPENDIX D. DATA REDUCTION D.1 Pressure Calculation The digits that are calculated from the voltage and current outputs are based on the equation: Digits = � � x 10 Period Digits= 1000 Equation 9 - Digits Calculation For example, a piezometer reading 8000 digits corresponds to a period of 354 µs and a frequency of 2828 Hz.
From → ↓ "H 2 O 'H 2 O mm H 2 0 m H 2 0 "HG mm HG mbar .036127 .43275 .0014223 1.4223 .49116 .019337 14.696 .014503 14.5039 .14503 145.03 "H 2 O 27.730 .039372 39.372 13.596 .53525 406.78 .40147 401.47 4.0147 4016.1...
D.3 Barometric Correction (required only on non-vented transducers) Since the standard piezometer is hermetically sealed and unvented, it responds to changes in atmospheric pressure. That being the case, corrections may be necessary, particularly for the sensitive, low-pressure models. For example, a barometric pressure change from 29 to 31 inches of mercury would result in ≈1 PSI of error (or ≈2.3 feet if monitoring water level in a well!).
The field value of C is calculated by inserting the initial zero reading into the polynomial equation with the pressure, P, set to zero. As of 8/2011, Geokon no longer includes the C coefficient on its calibration reports, ensuring that, to properly use the polynomial equation, users must calculate a C coefficient.
Referring to the example calibration report (Appendix E, Figure 36), note that the average zero reading (0 psi) is 9139 digits (8020-59 maximum digits) and the average maximum reading (100 psi) is 5691 digits (8020-59 minimum digits). These readings were recorded at an atmospheric pressure of 1001.4 mbar, corresponding to the conditions and altitude at the Geokon factory.