Function check
Result
6.4 Checking the fixation of the oxygen
Result
6.5 Checking the pressure reducer
18
EN
O
Box Rescue compact
2
12. Watch the needle on the contents gauge of the pressure
reducer for approx. 1 minute.
Requirement: The needle must remain still.
13. Set a locking button to 0 until the system is depressurized.
14. If a requirement is not fulfilled: contact
WEINMANN Emergency or a technician authorized by
WEINMANN Emergency.
The oxygen administrator has been checked.
administrator
1. Insert the oxygen administrator in the cover of the box.
2. Push the locking levers outwards until the locking mechanism
noticeably snaps into place.
Requirement: It must not be possible to move the oxygen
administrator.
3. If the locking levers do not function perfectly: replace the
equipment rail adapters.
The fixation has been checked.
Observe the instructions for use of the OXYWAY pressure reducer.