4.2 Changing the oxygen cylinder

Result
12
EN
O
Box Rescue compact
2
3. Open the oxygen cylinders.
The oxygen supply has been connected and the box is ready for
use.
1. Close the valve on the oxygen cylinder.
2. Open a dosage unit until the system is depressurized.
3. Remove the pressure hose from the pressure reducer.
4. Open the clamping belt on the oxygen cylinder.
5. Remove the oxygen cylinder from the box.
6. Disconnect the pressure reducer.
Risk of injury from flying particles of dust!
When opening the oxygen cylinder, flying particles of dust may
get into the eyes.
 Never aim the cylinder towards people when opening it.
7. Briefly open the valve of the new oxygen cylinder in order to
blow away any particles of dust.
8. Connect a pressure reducer to the valve of the new oxygen
cylinder using the knurled union nut.
9. Insert the oxygen cylinder in the box.
10. Tighten the clamping belt on the oxygen cylinder.
11. Reconnect the pressure hose.
The oxygen cylinder has been changed.