4
Preparation and operation
4.1 Connecting the oxygen supply to the
Risk of injury from highly compressed oxygen!
When combined with highly compressed oxygen, hydrocarbon
compounds can cause explosions.
Wash your hands before changing the oxygen cylinder.
Do not use any oil, grease, cleaning alcohols, hand cream, or
adhesive plasters.
Do not use tools to loosen or tighten the union nuts.
oxygen administrator
1. Connect the pressure hose's threaded connection to the free
threaded connection on the oxygen administrator.
2. Connect the pressure hose's self-sealing adapter to the
pressure reducer's Walther coupling.
Preparation and operation
O
Box Rescue compact
2
EN
11