Function check
6
Function check
6.1 Intervals
6.2 Checking the box
Result
16
EN
O
2
Carry out a function check at the following intervals:
•
Before each use
•
After each disassembly
•
Every 6 months
1. Carry out a visual inspection.
Requirement:
•
The corners and handle are undamaged.
•
The locks can be opened easily.
•
The hinges open smoothly and are correctly attached.
•
The foam is not impaired in its function.
•
The cover holder is undamaged and the hinge of the cover
holder opens smoothly.
•
The cover holder clicks into both positions (on the box and
on the cover) securely.
2. If one of the requirements is not fulfilled: replace defective
components or send the box to the manufacturer,
WEINMANN Emergency, for repair.
The box has been checked.
Box Rescue compact