Thermo Scientific Forma 8600 Series User Manual page 54

-40c and -86c ult chest freezer
Hide thumbs Also See for Forma 8600 Series:
Table of Contents

Advertisement

Certifications
Declaration of Conformity is available upon request.
Safety Specifications
Indoor Use Only
Altitude - up to 2,000 meters
Temperature - 5°C to 43°C
Humidity - Maximum RH 80% for temperatures up to 31°C, decreasing linearly to 50% RH at 40°C
Mains Supply Fluctuations - Mains supply voltage fluctuations not to exceed ±10% of the nominal voltage
Installation Category II
Pollution Degree 2
2
Class of Equipment I
Intended Use
This product is intended for use as a General Purpose Laboratory Freezer for storing samples or inventory between -50 and -86°C for units
rated at -86°C and between -10 and -40°C for units rated at -40°C.
This unit is not intended for use in an explosive environment, nor to be used for the storage of flammable inventory. This unit is not
intended for use in a Class II medical application as defined by Title 21 of the Federal Code of Regulations.
Fluorinated Greenhouse Gases
Compliant with REGULATION (EU) No 517/2014 OF THE EUROPEAN PARLIAMENT AND OF THE COUNCIL on fluorinated greenhouse gases.
This product contains fluorinated greenhouse gases in a hermetically sealed system. This product contains foam blown with fluorinated
greenhouse gas, R-245fa. If a leak in the sealed system is detected, the operator shall repair without undue delay
For ULT models rated 230V/50Hz, intended for import into the EU, the following sizes are designed for use with the following amounts of
fluorinated greenhouse gases:
For -86°C rated units
For -40°C rated units
Size (ft³)
Refrigerant
3
R-404a
12.7
R-404a
17
R-404a
20
R-404a
______________________________________________________________________
1 Installation category (overvoltage category) defines the level of transient overvoltage which the instrument is designed to withstand safely. It
depends on the nature of the electricity supply and its overvoltage protection means. For example, in CAT II which is the category used for
instruments in installations supplied from a supply comparable to public mains such as hospital and research laboratories and most industrial
laboratories, the expected transient overvoltage is 2500V for a 230V supply and 1500V for a 120V supply.
2
Pollution degree describes the amount of conductive pollution present in the operating environment. Pollution degree 2 assumes that
normally only non-conductive pollution such as dust occurs with the exception of occasional conductivity caused by condensation.
Thermo Scientific
1
1st Stage
Amount (kg)
0.425
0.539
0.595
0.652
GWP
3 922
3 922
3 922
3 922
Section 7
Specifications
.
8600 Series
7-5