Example
:SOURce?
Example
Chapter 5
Scanning Using External Triggers
In the following example, the trigger input is applied to the HP
E1300A/E1301A Mainframe's "Event In" port.
TRIG:SOUR EXT
SCAN (@100:115)
INIT
trigger externally
Scanning Using Bus Triggers
TRIG:SOUR BUS
SCAN (@100:115)
INIT
*TRG
TRIGger:SOURce?
Command returns either
sources
BUS
,
Query Trigger Source
TRIG:SOUR EXT
TRIG:SOUR?
HP E1351A/53A 16-Channel FET Multiplexer Command Reference
returns the current trigger source for the switchbox.
BUS
,
DBUS
,
DBUS
,
EXTernal
,
HOLD
Sets trigger source to external.
Sets channel list.
Starts scanning cycle.
Advances scan to next channel.
Sets interface (bus) triggering.
Sets channel list.
Starts scanning cycle.
Advances channel list.
EXT
,
HOLD
, or
IMM
for trigger
, or
IMMediate
, respectively.
Sets trigger source to external.
Query trigger source; returns
"EXT".
75