8.3 Phase

How to change the Phase ?
1. Press MENU to activate the Toolbar
2. Press → to select the Adjustments item
3. Press ↓ to Pull down the menu
4. Use ↑ or ↓ to select Phase
5. Press ENTER
A sliderbox is displayed
6. Use ←or →, the numeric keys on the remote, or the keypad to change the Phase.

8.4 Gamma

Gamma
Signal-Light transfer characteristic of a display.
Gamma adjustment
In the MGP10 the Gamma adjustment remains only adjustable in the Standard Display Function i.e. only if used for projection without
any constraint with with respect to consistency and conformance. The adjustment allows an adjustment in a range from 1.6 to 2.8.
A gamma of 2.2 is considered as the typical value for video signals.
When a Display Function is chosen to be DICOM, DIN or Custom then the Gamma function or more generally the signal to luminance
characteristic is at the same time determined.
How to change the Gamma ?
1. Press MENU to activate the Toolbar
2. Press → to select the Adjustments item
3. Press ↓ to Pull down the menu
4. Use ↑ or ↓ to select Gamma
5. Press ENTER
A sliderbox is displayed.
6. Use ←or →, the numeric keys on the remote, or the keypad to change the Gamma.
Gamma adjustment is only possible if the Display function is Standard.
In the other modes (DICOM, DIN, Custom ) the Gamma adjustment is disabled since the signal-light function
is determined by the Display Function (DICOM, DIN, Custom )

8.5 Custom color temperature adjustment

How to edit the custom color temperature ?
1. Press MENU to activate the Toolbar
2. Press → to select the Adjustments item
3. Press ↓ to Pull down the menu
4. Use ↑ or ↓ to select Color temperature
5. Press → to open the Color temperature item and select Edit RGB (image 8-4)
6. Press ENTER
A sliderbox for the Red custom setting is displayed as well as a wizard textbox in the lower part of the screen. (image 8-5,
image 8-6)
7. Use ←or →, the numeric keys on the remote, or the keypad to change the Color temperature.
8. Follow the instructions on the wizard textbox.
R5976741 MGP 15 MEDICAL GRADE PROJECTOR 28/04/2004