Use of incorrect cables may result in the device causing radio frequency interference.
Electrostatic discharges environment & recommendations
In order to reduce electrostatic discharge interference, a charge dissipative floor should
be installed to prevent charge accumulation.
• The dissipative floor material must be connected to the system reference ground, if
applicable.
• Relative humidity must be maintained above 30 percent.
Stacked components & equipment
WARNING! Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation. If such use is
necessary, this equipment and the other equipment should be observed to verify
that they are operating normally.
Interference may occur in the vicinity of equipment marked with the following symbol:
WARNING! Portable RF communications equipment (including peripherals
such as antenna cables and external antennas) should be used no closer than
30 cm (12 inches) to any part of the FS Ergo, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could
result."