orienting or relocating the device. Suggested actions according to the table RF immunity
of non-life-support equipment or system IEC 60601-1-2 ed.4 .
The device is suitable for use in both professional healthcare (hospitals/large clinics)
facility environment and home healthcare (clinics in domestic establishments and those
directly connected to the public low-voltage power supply) environment.
Exceptions for professional healthcare facility environment, not to be used or installed:
• near active HF SURGICAL EQUIPMENT
• near the RF shielded room of an ME SYSTEM for magnetic resonance imaging, where
the intensity of EM DISTURBANCES is high.
CAUTION! Using non-USB compliant cables or hubs, or exceeding the maximum
count of USB hub devices for extending the distance, can degrade the immunity
of the sensor to electromagnetic fields or increase the emission of electromagnetic
fields from the sensor.
NOTICE! Portable/mobile radio frequency communications equipment can affect
the function of the sensor as well as any other electronic medical equipment.
Table 1: Electromagnetic emissions IEC 60601-1-2 Edition 4.0 2014-02
FS Ergo is suitable for use in the specified electromagnetic environment.
The purchaser or user of FS Ergo should assure that it is used in an
electromagnetic environment as described below:
Emissions Test
Radio-Frequency
Emissions CISPR11
Radio-Frequency
Emissions CISPR11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations/
flicker emissions
IEC61000-3-3
FS Ergo Sensor
Compliance Electromagnetic Environment
Group 1
FS Ergo uses RF energy only for its internal
function. Therefore, the RF emission is very low
and not likely to cause any interference in nearby
electronic equipment.
Class B
FS Ergo is suitable for use in both professional
healthcare (hospitals/large clinics) facility
environment and home healthcare (clinics in
IEC61000-3-2
domestic establishments and those directly
connected to the public low-voltage power supply)
Class B
environment.
Complies
Exceptions for professional healthcare facility
environment: Not to be used or installed near
active HF SURGICAL EQUIPMENT and the RF
shielded room of an ME SYSTEM for magnetic
resonance imaging, where the intensity of EM
DISTURBANCES is high.
7 Technical data
33