7 Technical data
Table 2: Electromagnetic immunity IEC 60601-1-2 Edition 4.0 2014-02
FS Ergo is suitable for use in the specified electromagnetic environment.
The purchaser or user of the FS Ergo should assure that it is used in an
electromagnetic environment as described below:
Immunity Test IEC60601-1-2
Electrostatic
discharge (ESD)
IEC61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC61000-4-5
Voltage
dips, short
interruptions
and voltage
variations on
power supply
input lines
IEC61000-4-11
Power
frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
NOTICE! U
34
Compliance
Test Level
Level
± 8 kV for
± 8 kV for
contact
contact
discharge
discharge
±2, 4, 8, 15 kV
±2, 4, 8,15 kV
for air discharge
for air discharge
±2 kV for power
±2 kV for power
supply lines
supply lines
(100 kHz)
±1 kV for input/
±1 kV for input/
output lines
output lines
(100kHz)
± 0.5, 1 kV
± 0.5, 1 kV
differential
differential
mode
mode
± 0.5, 1, 2 kV
± 0.5, 1, 2 kV
common mode
common mode
-0 % U mains;
-0 % U mains;
0,5 cycle At 0°,
0,5 cycle At 0°,
45°, 90°, 135°,
45°, 90°, 135°,
180°, 225°,
180°, 225°,
270° and 315°
270° and 315°
-0 % U mains;
-0 % U mains;
1 cycle, At 0°
1 cycle, At 0°
-70 % U mains;
-70 % U mains;
25/30 cycles At
25/30 cycles At
-0 % U mains;
-0 % U mains;
250/300 cycle
250/300 cycle
At 0°
At 0°
30A/m
30 A/m
is the a.c. mains voltage prior to application of the test level.
T
Electromagnetic Environment
Floors are wood, concrete, or
ceramic tile, or floors are covered
with synthetic material and the
relative humidity is at least 30
percent.
Mains power quality is that of a
typical commercial and/or hospital
environment
Mains power quality is that of a
typical commercial and/or hospital
environment.
Mains power quality is that of
a typical commercial and/or
hospital environment. If the user
of FS Ergo requires continued
operation during power mains
interruptions, it is recommended
that FS Ergo be powered from
an uninterruptible power supply.
Power frequency magnetic fields
are at levels characteristic of
a typical location in a typical
commercial and/ or hospital
environment.
The device shall not be used
closer than 15cm to sources of
50/60Hz magnetic field.
FS Ergo Sensor