Although new parameter values are stored immediately when the
command is issued, they do not take effect until the sensor enters
OPERATE MODE.
COMMAND
DESCRIPTION
SET01nn
Set OUTPUT MODE
01
Enable voltage output
02
Enable RS-232
04
Enable RS-485 full duplex
08
Enable RS-485 half duplex
16
Enable RS-485 output only
SET02nn
Set OUTPUT FORMAT
01
RMYT (default)
02
ASCII
03
ASCII POLLED
04
NMEA
05
SDI-12 CONTINUOUS
06
SDI-12 STANDBY
SET03nn
Set BAUD RATE
12
1200
48
4800
96
9600
38
38400
SET04nn
Set ASCII WS UNITS
01
MPH
02
KNOTS
03
KMPH
04
M/S
SET05c
ASCII character sensor address (0-9, A-Z, a-z)
SET06nnn
Wind speed threshold for polar output (cm/s)
SET07nnnnn
Wind speed scale (nnnnn/10000)
SET08nnnnn
Direction offset (±nnnnn degrees x 10)
SET09nn
Damping factor
SET10nnn
Output delay (0-999 x 8.192 milliseconds)
SET11nn
Direction VOUT (36=0-360, 54=0-540 degrees)
SET12nn
Sample count (00=auto or 03-31)
SET13nn
ASCII and Vout format (00=polar, 01=UV)
SET14nn
Vout Status Code (1=none, 2=0V, 3=FS)
CALnn
Force Vout (00=0V, 01=5V)
XX
Go to OPERATE MODE
RPT
Report parameter settings
??
Help list

6.4 COMMAND DETAILS

6.4.1 SET01nn SET OUTPUT MODE
This parameter independently enables and disables voltage output,
RS-232, and RS-485 connections. For example, to enable voltage
output, RS-232, and RS-485 full duplex, add the code for each one:
01 + 02 + 04 = 07, SET0107. To enable only RS-232: SET0102. If
RS-485 is used, choose only one RS-485 option (04, 08, or 16). See
section 6.3 for codes.
To reduce power consumption, enable only those outputs that are
needed. When SDI-12 format is selected with the SET02nn com-
mand, voltage output, RS-232, and RS-485 connections are disabled
and SDI-12 is enabled.
6.4.2 SET02nn SET OUTPUT FORMAT
This parameter determines the serial output data format. See section
5.0 for operational details. When selecting some formats, parameters
are automatically changed as listed below:
FORMAT
CODE
01 RMYT
02 ASCII
03 ASCII POLLED
04 NMEA
05 SDI-12 CONT
06 SDI-12 STANDBY
Both SDI-12 settings automatically disable voltage output, RS-232,
and RS-485 connections. With either POLLED ASCII or SDI-12
STANDBY, set the Damping Factor to 0 (see section 6.4.9).
6.4.3 SET03nn SET BAUD RATE
All RS-232 and RS-485 communication use the baud rate set with
this parameter. Power consumption is affected by this setting since
it influences communication duration.
6.4.4 SET04nn SET ASCII WIND SPEED UNITS
Sets wind speed units for ASCII, ASCII POLLED, and SDI-12 polar
outputs only.
6.4.5 SET05c SET POLL ADDRESS CHARACTER
Sets the sensor address for ASCII POLLED and SDI-12 formats. The
default is '0' (ASCII 48).
6.4.6 SET06nn SET WIND SPEED THRESHOLD
Sets the wind speed threshold for polar outputs (wind speed and
direction) to suppress scattered wind direction indications in calm
air. Winds below the threshold are reported as zero, wind direction
is held at last value when wind speed was above threshold. Set
threshold in centimeters per second (m/s x 100). The default setting
is 25 cm/s (0.25 m/s, 0.56 mph).
6.4.7 SET07nnnnn SET WIND SPEED MULTIPLIER
All wind speed measurements are multiplied by this parameter. The
default value is 10000 for a multiplier of 1.0000. This may be used to
individually calibrate each sensor for improved accuracy.
6.4.8 SET08nnnnn SET WIND DIRECTION OFFSET
Use this parameter to add or subtract a wind direction offset. Value
is degrees x 10 and may be positive or negative. Wind direction is
always re-scaled to a 0-360 range after offset is applied. The default
value is 00000.
6.4.9 SET09nn SET DAMPING FACTOR
Wind measurement outputs are damped using the following formula:
S
= [(d-1) * S
damped
damped
where:
S
= New or last damped wind speed
damped
S
= New wind speed speed
sample
d
= Damping factor
Page 5
ADDITIONAL
PARAMETERS CHANGED
BAUD = 9600
OUTPUT MODE = 07
THRESHOLD = 25 cm/s
OUTPUT DELAY = 01
SAMPLE COUNT = 31
None
BAUD = 4800
OUTPUT MODE = 19
THRESHOLD = 25 cm/s
OUTPUT DELAY = 122
SAMPLE COUNT = 31
1200 baud