Operation
3.12 Change the user-selected temperature
compensation factor
Select the User mode to change the temperature compensation factor. This feature
does not apply to pH or ORP. For user mode information, (see
page
12).
Select the User mode (see
1.
Push the CAL/MCLR key.
2.
Push UP/MS or MR/DOWN to adjust the temperature compensation factor
3.
from 0-9.99%/
Push the CAL/MCLR key twice to skip calibration adjustment and accept the
4.
new temperature compensation (three times if in TDS or MIN/SALT mode).
Measure samples with the new temperature compensation factor.
5.

3.12.1 Disable temperature compensation

Select the User mode (see
1.
Push the CAL/MCLR key. Hold the MR/DOWN key until the temperature
2.
compensation shows .00%/
Push the CAL/MCLR key twice (three times for TDS or MIN/SALT).
3.
Temperature compensation is now disabled (=0) for measurements in the User
4.
mode.

3.13 Change the user-selected conductivity/TDS ratio

Select the User mode to change a custom conductivity/TDS conversion ratio in the
range of 0.20 to 7.99.
To determine the conversion ratio for a custom solution of a known TDS ppm value,
measure the solution conductivity at 25 ºC with the MP Series meter and divide the
20
section 3.11 on page
C
º
.
section 3.11 on page
C.
º
section 2.2.3 on
19).
19).