Replacing the Power Cord

Hard-wired Cords on Low Voltage Units

Tools: Heyco® strain relief hand pliers
Remove the configuration plate as described on the previous page.
1.
Use strain relief pliers to squeeze the strain relief on the power cord and pull it free of
2.
the configuration plate.
Clip the power cord wire ties.
3.
Remove the two hex nuts holding the power cord to the line filter: the hex nuts are
4.
behind the configuration plate, down and to the right.
Pull the power cord out of the configuration plate.
5.
Feed the new cord into the configuration plate and complete the procedure by
6.
following steps 1 through 4 in reverse order.
Twist the power leads before connecting.
a.
Connect the white wire to the bottom connector on the line filter.
b.
If no other service is required, replace the treadmill hood as described in
7.
page
5-4.

Detachable Cords on High Voltage Units

Unplug the power cord from the connector on the configuration plate and plug in the new
cord.
TM 55/65 Treadmill 4.0
Replacing the Power Cord
step 9 on
70-00298-01 A
5-7