Warnings and Cautions

Before operating any treadmill, ensure the following warnings and cautions are understood
and followed.
WARNING!
The treadmill must be on an appropriate, dedicated electrical circuit with a power rating that
meets the electrical specification on the treadmill serial number label. Nothing else should
be connected to the circuit.
WARNING!
As long as the treadmill is plugged into a powered outlet and the treadmill circuit breaker is
set to ON, the treadmill is receiving power, even when the monitoring system is turned off.
Do not place hands beneath the treadmill while it is plugged in.
Caution
Read this manual in full before operating the treadmill.
Caution
Before each use of this equipment, check the power receptacle for signs of damage. Do not
operate the equipment if the integrity of these items is in question.
Caution
Be sure the connection for the optional Emergency Off switch is secure. Test the switch to
verify proper operation. Do not operate the equipment if the integrity of these items is in
question.
Caution
Keep the treadmill area clear. Maintain a minimum open space of 1.5 feet (0.5 meter) on each
side and 6 feet (2 meters) at the rear.
Caution
The Cardiac Stress Treadmill should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the Cardiac Stress Treadmill should be
observed to verify normal operation in the configuration in which it will be used.
Caution
Do not operate this equipment in the presence of flammable anesthetic mixtures.
Caution
Regularly inspect cables and treadmill belts for wear or damage. Do not operate the
equipment if the integrity of these items is in question.
TM 55/65 Treadmill 4.0
Warnings and Cautions
70-00298-01 A