Capture Page - Mantracourt FSU-SSBD User Manual

Fast usb strain gauge/load cell digitiser module
Table of Contents

Advertisement

Capture Page

Click on the
icon to open the Capture page.
This page helps you to see what is happening in real-time to the input of the module. All the other pages offer you
the options of Capture All to capture and review everything recorded so far. This page allows you to capture shorter
durations and to see the state of the input for those times when you can identify when to capture the input data.
View Window
Select the viewing window into the real-time data by selecting between 1 and 30 seconds using the slider. The input
data is displayed on the chart.
Transient Enhance
Selecting this option will enhance the display to better identify fast transient peaks. Because of the way a large
amount of data is subsampled to enable it to fit on the chart, fast peaks may not appear or may flash in and out of
visibility as the data flows past the chart. This feature will have the side effect of creating a slightly fuzzy edge on
slower moving data. This does not affect the way data is recorded, only the way it is shown on the chart. When used
in conjunction with the Show Min Max feature this can help stabilize the view of data.
Show Min Max
By default the chart auto scales as the data moves across it which means that if a transient occurs the chart will
rescale but once that transient leaves the viewport the chart will autoscale again based on the lower magnitude data
that follows. By selecting this feature the chart scaling will be locked to the minimum and maximum values detected
while the feature is active.
Also, when activated, the values of the detected min and max along with the difference between the two are
displayed in the lower right of the window in the Help pane.