System Zero

System Zero is used to remove permanent values from an installed system such as the weight of a vessel or weighing
platform to give a zero Gross value.
The value shown on this page is subtracted from the measured value.
You can either type a value in manually or click the Zero Now button to automatically fill in the value to give a zero
reading based on the current input.
This feature can be hidden on the
Lock
Page.

Auto Calibration

The automatic calibration requires that you apply two known (ascending) inputs in sequence. Just enter the required
engineering unit value for the two inputs in the fields marked 1 and 2.
Next, apply the low input and allow a settle time. Click the 'Acquire' button then apply the higher input. After
another settle time click the second 'Acquire' button and the calibration is complete.
13
Mantracourt Electronics Limited – FSU User Manual