Lock Page - Mantracourt FSU-SSBD User Manual

Fast usb strain gauge/load cell digitiser module
Table of Contents

Advertisement

Lock Page

Click on the
icon to open the Lock page.
No data is recorded while on this page.
This page will allow the locking of certain functions. Once any item has been changed from Show to Hide or Allow to
Lock then subsequent attempts to open this page (And thus show the hidden functions again) will result in the user
being asked for a Passcode.
The Passcode entered must match the one entered on this page to grant access.
It is important to remember the Passcode before leaving this page after any of the options have
been set to Hide. You will need to contact Mantracourt to aid in Passcode recovery should the
Passcode be forgotten!
Calibration Page – Allows you to either Allow or Lock access to the calibration page.
System Zero Function – Locking this function will also force locking of the Calibration Page. This is useful if the
Calibration Page has been locked but you want to still offer the ability to set the System Zero. Just set this to Allow
after setting the Calibration Page to lock.
Display Units Change - Hiding this function will also force locking of the Calibration Page. This is useful if you want
to lock access to the Calibration Page but allow the user to change the displayed engineering units. Just set this to
Show after setting the Calibration Page to lock.
11
Mantracourt Electronics Limited – FSU User Manual