Omron JPN1 Instruction Manual

Omron JPN1 Instruction Manual

Automatic blood pressure monitor
Hide thumbs Also See for JPN1:

Advertisement

Quick Links

1. Overview
Main unit:
A
B
Automatic Blood
F
Pressure Monitor
C
E
Model JPN1
D
D
Instruction Manual
A. Display
B. START/STOP button
1883264-4A
C. Memory button
D. Up/Down (
) buttons
E. Buzzer and Date/Time
setting (
) button
Introduction
Thank you for purchasing the OMRON JPN1 Automatic Blood Pressure
Monitor.
Arm cuff:
The OMRON JPN1 is a compact, fully automatic blood pressure monitor,
I
J
K
operating on the oscillometric principle. It measures your blood pressure and
pulse rate simply and quickly. For comfortable controlled inflation without the
need of pressure pre-setting or re-inflation the devices uses its advanced
"IntelliSense" technology.
Intended Use
This product is designed to measure the blood pressure and pulse rate of
L
people within the range of the designated arm cuff, following the instructions in
this instruction manual.
It is mainly designed for general household use. Please read the Important
Safety Information in this instruction manual before using the unit.
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure, CONSULT YOUR
DOCTOR.
Display:
M
Important Safety Information
N
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
O
or arteriosclerosis.
P
Please read this section carefully before using the unit.
Q
Warning:
• Indicates a potentially hazardous situation which, if not avoided, could result
R
in death or serious injury.
S
(General Usage)
T
• Always consult your doctor. Self-diagnosis of measurement results and
self-treatment are dangerous.
M. Systolic blood pressure
• People with severe blood flow problems, or blood disorders, should consult a
N. Diastolic blood pressure
doctor before using the unit, as cuff inflation can cause internal bleeding.
O Heartbeat symbol
(AC Adapter Usage)
(Flashes during measurement)
• Never plug in or unplug the power cord from the electric outlet with wet
P. Memory symbol
hands.
Q. Deflation symbol
(Battery Usage)
R. Average value symbol
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
water. Consult a doctor immediately.
Caution:
• Indicates a potentially hazardous situation which, if not avoided, may result in
2. Preparation
minor or moderate injury to the user or patient or damage to the equipment or
other property.
2.1 Installing/Replacing the Batteries
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express
1.
Remove the battery cover.
their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
2.
• Do not disassemble the unit or arm cuff.
Insert four "AA" batteries
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result
as indicated in the battery
in incorrect measurement results.
compartment and then
• Make sure that the air tube is not wrapped around other parts of your body
replace the battery cover.
when taking measurements. This could result in injury when the air pressure
in the air tube is increased.
Notes:
• Do not leave the cuff wrapped on the arm if taking measurements during the
night. This could result in injury.
• If the battery low symbol (
• Do not inflate the arm cuff over 299 mmHg.
then replace all batteries at the same time.
• Do not use a mobile phone or other devices that emit electromagnetic fields,
• The measurement values continue to be stored in memory even after
the batteries are replaced.
near the unit. This may result in incorrect operation of the unit.
• Do not operate the unit in a moving vehicle (car, airplane).
Disposal of used batteries should be carried out in accordance with the
• To inflate the cuff manually, refer to Section 3.3. If the cuff is over inflated, it
national regulations for the disposal of batteries.
can cause internal bleeding.
(AC Adapter Usage)
• Use only the original AC adapter (optional) designed for this unit. Use of
2.2 Setting the Buzzer/ Date and Time
unsupported adapters may damage and/or may be hazardous to the unit.
• Plug the AC adapter into the appropriate voltage outlet. Do not use a
1.
Press the
button.
multiple-tap.
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off
2.
the power and unplug the power cord immediately.
Set the monitor to the correct date and time before
(Battery Usage)
taking a measurement for the first time.
• If battery fluid should get on your skin or clothing, immediately rinse with
plenty of clean water.
• Use only four "AA" alkaline or manganese batteries with this unit. Do not use
1)
other types of batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
2)
• Replace old batteries with new ones immediately. Replace all four batteries at
the same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
Year
Month Day
• Do not take measurements after bathing, drinking alcohol or caffeine,
smoking, exercising or eating.
• Do not forcibly bend the arm cuff or bend the air tube excessively.
3.
Press the START/STOP button to store the setting.
• When removing the air tube, pull on the air plug at the connection with the
Notes:
main unit not the tube itself.
• If the batteries have been removed for 30 seconds or more, the date
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not wash the arm cuff or immerse it in water.
and time setting will need to be reset.
• Read and follow the "Important information regarding Electro Magnetic
• If the date and time are not set, "-:--" appears during or after
measurement.
Compatibility (EMC)" in the Technical Data Section.
• Read and follow the "Correct Disposal of This Product" in the Technical Data
Section when disposing of the device and any used accessories or optional
parts.
3. Using the Unit
3.1 Applying the Arm Cuff
G
Remove tight-fitting clothing or tight rolled up sleeve from your
upper arm.
Do not place the cuff over thick clothes.
H
1.
Insert the air plug into the air jack.
F. Air Jack
G. Battery compartment
H. AC adapter jack (for optional AC
adapter)
2.
Put your arm through the cuff loop.
I. Arm cuff (Medium cuff: arm
circumference 22-32cm)
J. Marker
K. Air tube
3.
L. Air plug
Position the arm correctly.
U
The bottom edge of the cuff should be 1 to 2 cm above the elbow.
Marker (arrow under tube) is centred on the middle of your inner arm.
V
Close the fabric fastener FIRMLY.
1 to 2 cm
W
X
S. Movement error symbol
Notes:
T. Date/Time display
• When you take a measurement on the right
U. Blood pressure level indicator
V. Pulse display
arm, air tube will be at the side of your
elbow. Be careful not to rest your arm on
W. Battery low symbol
the air tube.
X. Irregular heartbeat symbol
• The blood pressure can differ between the right arm and the left arm,
and therefore also the measured blood pressure values can be
different. OMRON recommends to always use the same arm for
measurement. If the values between the two arms differ substantially,
please check with your doctor which arm to use for your
measurement.
3.2 How to Sit Correctly
To take a measurement, you need to be
relaxed and comfortably seated, under
comfortable room temperature. No eating,
smoking or exercising 30 minutes before
taking a measurement.
• Sit on a chair with your feet flat on the floor.
) appears on the display, turn the unit off
• Sit upright with your back straight.
• The cuff should be at the same level as your
heart.
3.3 Taking a Reading
Notes:
Push
or
buttons to
• To cancel a measurement, press the START/STOP button to turn off
change.
the unit and to release the air in the arm cuff.
Push the
button to confirm.
• Remain still while taking a measurement.
Next setting appears.
1.
Press the START/STOP button.
The cuff will start to inflate automatically.
START
INFLATE
DEFLATE
END
COMPLETE
Hour Minute
If your systolic pressure is more than
220 mmHg
After the cuff starts to inflate, press and hold the
START/STOP button until the monitor inflates 30 to 40 mmHg
higher than your expected systolic pressure.
Notes:
• The monitor will not inflate above 299 mmHg.
• Do not apply more pressure than necessary.
Note: Wait 2-3 minutes before taking another blood pressure
4. Troubleshooting and Maintenance
measurement. Waiting between readings allows the arteries to
4.1 The Icons and Error Messages
return to the condition prior to taking the blood pressure
measurement.
Error Display
Cause
Warning:
Irregular heartbeats are
Self-diagnosis of measured results and treatment are dangerous.
detected.
Please follow the instructions of your doctor.
Movement during
2.
Remove the arm cuff.
measurement.
3.
Press the START/STOP button to turn off the monitor.
The monitor automatically stores the measurement in its
Blink
The batteries are low.
memory.
It will automatically turn off after five minutes.
The batteries are
Important:
Lit
exhausted.
• Recent research suggests that the following
values can be used as a guide to high blood
pressure for measurements taken at home.
Cuff is under inflated.
Movement during
measurement.
Systolic Blood Pressure
Above 135 mmHg
Diastolic Blood Pressure
Above 85 mmHg
Air plug disconnected.
This criteria is for home blood pressure measurement.
• Your blood pressure monitor includes an irregular
Arm cuff not applied
heartbeat feature. Irregular heartbeats can influence
correctly.
the results of the measurement. The irregular
heartbeat algorithm automatically determines if the
Clothing is interfering with
measurement is usable or needs to be repeated. If the
the arm cuff.
measurement results are affected by irregular
heartbeats but the result is valid, the result is shown
together with the irregular heartbeat symbol (
). If the
Air is leaking from the arm
irregular heartbeats cause the measurement to be
cuff.
invalid, no result is shown. If the irregular heartbeat
symbol (
) is shown after you have taken a measurement, repeat
The arm cuff was inflated
the measurement. If the irregular heartbeat symbol (
) is shown
above 299 mmHg when
frequently, please make your doctor aware of it.
inflating the cuff manually.
3.4 Using the Memory Function
Device error.
The monitor automatically stores the result up to 60 sets. It can also
calculate an average reading based on the measurements from the last
three readings taken within 10 minutes. If there are only two readings in
memory for that period, the average will be based on two readings. If
Note: The irregular heartbeat symbol (
there is one reading in memory for that period, the average will be
messages.
based on one reading.
Notes:
4.2 Troubleshooting
• If the memory is full, the monitor will delete the oldest readings.
• When viewing the reading taken without setting the date and time,
"-:--" is displayed instead of the date and time.
Problem
Cause
Average Value
Previous Reading
Arm cuff not applied
correctly.
Alternating
date/time display
The reading is
Movement or talking
extremely low (or
during measurement.
high).
Old
Recent
Clothing is interfering
with the arm cuff.
View the Average
View the Previous
View the Readings
Value
Reading
Stored in Memory
• If the previous
• The Memory number
The air tube is not
reading was taken
appears for a second
securely connected
without setting the
before the pulse rate
into the main unit.
Arm cuff pressure
date and time, the
is displayed. The
does not rise.
average value is not
newest set is
calculated.
numbered "1".
Air is leaking from the
arm cuff.
To delete all the values stored in memory
Arm cuff deflates too
The arm cuff is loose.
When the memory symbol (
) appears, first press the Memory
soon.
button. Then while holding it down, press the START/STOP button
simultaneously for about 2-3 seconds.
Cannot measure or
The arm cuff has not
readings are too low
been inflated
or too high.
sufficiently.
The batteries are
empty.
Nothing happens
when you press the
Note: You cannot partially delete the stored readings.
The batteries have
buttons.
been inserted
incorrectly.
• Press the START/STOP button and repeat
measurement.
• If the problem continues, try replacing the
Other problems.
batteries with new ones.
If this still does not solve the problem, contact
your local OMRON representative.
4.3 Maintenance
To protect your unit from damage, please observe the following:
• Do not subject the main unit and the cuff to extreme
Remedy
temperatures, humidity, moisture or direct sunlight.
• Do not fold the cuff or tubing tightly.
Remove the arm cuff. Wait
2-3 minutes and then take
• Do not inflate the arm cuff over 299 mmHg.
another measurement.
• Do not disassemble the unit.
Repeat the steps in section
• Do not subject the unit to strong shocks or vibrations
3.3. If this error continues to
(for example, dropping the unit on the floor).
appear, contact your doctor.
• Do not use volatile liquids to clean the main unit.
Carefully read and repeat
• Do not wash the arm cuff or immerse it in water.
the steps in section 3.3.
• Do not use petrol, thinners or similar solvents to clean
the arm cuff.
You should replace them
• Do not carry out repairs of any kind yourself. If a defect
with new ones ahead of
time.
occurs, consult your local OMRON representative.
Refer to section 2.1.
You should replace them
with new ones at once.
Refer to section 2.1.
Carefully read and repeat
the steps listed under
section 3.3.
Repeat measurement.
Remain still and do not talk
during measurement.
Refer to section 3.3.
Insert the air plug securely.
Refer to section 3.1.
Apply the arm cuff correctly.
Refer to section 3.1.
• The unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and soap to clean the arm
Remove any clothing
cuff.
interfering with the arm cuff.
Refer to section 3.1.
Calibration and Service
• The accuracy of this blood pressure monitor has been
Replace cuff with new one.
carefully tested and is designed for a long service life.
Refer to Chapter 5.
• It is generally recommended to have the unit inspected
every two years to ensure correct functioning and
Do not inflate the arm cuff
above 299 mmHg.
accuracy. Please consult your local OMRON
Refer to section 3.3.
representative.
Contact your local OMRON
representative.
4.4 Storage
) may also be displayed with error
1.
Unplug the air tube from the air connector.
2.
Gently fold the air tube into the arm cuff.
Remedy
Note:
Do not bend the air tube excessively.
Apply the arm cuff
correctly.
Refer to section 3.1.
Remain still and do not
talk during
measurement.
Refer to section 3.3.
Do not store the unit in the following situations:
Remove any clothing
• If the unit is wet.
interfering with the arm
• Locations exposed to extreme temperatures, humidity, direct sunlight,
cuff.
dust or corrosive vapours.
Refer to section 3.2.
• Locations exposed to vibrations, shocks or where it will be at an
angle.
Make sure that the air
tube is connected
securely.
Refer to section 3.1.
Replace the arm cuff
5. Optional Parts
with a new one.
Refer to Chapter 5.
Medium Arm Cuff
Large Arm Cuff
AC Adapter
Apply the cuff correctly
Arm circumference
Arm circumference
so that it is firmly
22 - 32 cm
32 - 42 cm
wrapped around the arm.
Refer to section 3.1.
Inflate the cuff so that it
is 30 to 40 mmHg above
your previous
measurement result.
Note: Please check with your local OMRON representatives for the
Refer to section 3.3.
appropriate optional part models.
Replace the batteries
with new ones.
Refer to section 2.1.
Insert the batteries with
the correct (+/-) polarity.
Refer to section 2.1.
Using the Optional AC Adapter
1.
Insert the AC adapter plug
into the AC adapter jack on
the rear side of the main
unit.
2.
Plug the AC adapter into an
electrical outlet.
To disconnect the AC adapter, unplug the AC adapter from the
electrical outlet first and then remove the AC adapter plug from the
main unit.
HEM-7200-AP3_A_M03_120823.pdf
6. Technical Data
Product Description
Automatic Blood Pressure Monitor
Model
JPN1
Display
LCD Digital Display
Measurement Method
Oscillometric method
Measurement Range
Pressure: 0 to 299 mmHg
Pulse: 40 to 180/min.
Accuracy
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Inflation
Fuzzy-logic controlled by electric pump
Deflation
Automatic pressure release valve
Memory
60 Measurements with date and time
Rating
DC6V 4W
Power Source
4 "AA" batteries 1.5V
or AC adapter (optional, INPUT AC100-240V
50/60Hz 0.12A)
Battery Life
Approx. 1500 measurements with new alkaline
batteries at a room temperature of 23°C
Applied Part
= Type B
Protection Against
Internally powered ME equipment (When using only
Electric Shock
the batteries)
= Class II ME equipment (Optional AC adapter)
Operating
+10 to +40°C / Maximum: 30 to 90% RH
temperature/ Humidity
Storage temperature/
-20 to +60°C / Maximum: 10 to 95% RH
Humidity/ Air pressure
/ 700-1060 hPa
Console Weight
Approx. 340g without batteries
Cuff Weight
Approx. 120g
Outer Dimensions
Approx. 123 (w) mm × 85 (h) mm × 141(l) mm
Cuff Dimensions
Approx. 146 mm × 446 mm (Medium cuff: arm
circumference 22 to 32 cm)
Cuff Material
Nylon and polyester
Package Content
Main unit, cuff, instruction manual, battery set
Note: Subject to technical modification without prior notice.
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
• This blood pressure monitor is designed according to the European Standard
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements
and Part 3: Supplementary requirements for electromechanical blood
pressure measuring systems.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC's and mobile
(cellular) telephones, medical devices in use may be susceptible to
electromagnetic interference from other devices. Electromagnetic interference
may result in incorrect operation of the medical device and create a potentially
unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)
with the aim to prevent unsafe product situations, the EN60601-1-2:2007
standard has been implemented. This standard defines the levels of immunity to
electromagnetic interferences as well as maximum levels of electromagnetic
emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This may
result in incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct operation
of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at
OMRON HEALTHCARE EUROPE at the address mentioned in this instruction
manual.
Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it
should not be disposed of, with other household wastes at the end
of its working life. To prevent possible harm to the environment or
human health from uncontrolled waste disposal, please separate
this from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they can
take this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions
of the purchase contract. This product should not be mixed with other commercial
wastes for disposal.
This product does not contain any hazardous substances.
Manufacturer
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto
617-0002 JAPAN
EU-representative
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp
THE NETHERLANDS
www.omron-healthcare.com
Production facility
OMRON HEALTHCARE Co., Ltd.
Matsusaka Factory
Mie, JAPAN
Asia Pacific HQ
OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road, #05-05/08
Alexandra Technopark, Singapore 119967
www.omron-healthcare.com.sg
Made in Japan