Omron JPN2 Instruction Manual
Omron JPN2 Instruction Manual

Omron JPN2 Instruction Manual

Automatic blood pressure monitor

Advertisement

Quick Links

1. Overview
Main unit:
A
Automatic
Blood Pressure Monitor
Model
JPN2
Instruction Manual
B
C
A. Display
B. START/STOP button
8500188-9A
C. M (Memory) button
Display:
Introduction
G
Thank you for purchasing the OMRON JPN2 Automatic Blood Pressure
H
Monitor.
I
The JPN2 is a fully automatic blood pressure monitor, operating on the
J
oscillometric principle. It measures your blood pressure and pulse rate simply
and quickly. For comfortable controlled inflation without the need of pressure
G.
Systolic blood pressure
pre-setting or re-inflation the device uses its advanced "IntelliSense" technol-
H.
Diastolic blood pressure
ogy.
I.
Memory symbol
(Displayed when viewing
Intended Use
values stored in memory)
This product is designed to measure the blood pressure and pulse rate of
J.
Irregular heartbeat symbol
people within the range of the designated arm cuff, following the instructions in
K.
Heartbeat symbol
this instruction manual.
1.
Flashes during measurement
It is mainly designed for general household use. Please read the Important
2.
If flashing after measurement
Safety Information in this instruction manual before using the unit.
complete indicates blood pressure
Please read this instruction manual thoroughly before using the unit.
out of recommended range
Please keep for future reference. For specific information about your own
blood pressure, CONSULT YOUR DOCTOR.
Arm cuff:
O
P
Important Safety Information
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
or arteriosclerosis.
Q
Please read this section carefully before using the unit.
Warning: Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
2. Preparation
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
2.1 Installing/Replacing the Batteries
self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a
1.
doctor before using the unit, as cuff inflation can cause internal bleeding.
Turn the main unit upside down.
(AC Adapter (Optional) Usage)
2.
Slide the battery cover in the direction
• Never plug in or unplug the power cord from the electric outlet with wet
of the arrow while pressing the ribbed
hands.
part of the cover.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
3.
Install or replace four "AAA" size
water. Consult a doctor immediately.
batteries so that the + (positive)
Caution: Indicates a potentially hazardous situation which, if not avoided,
and - (negative) polarities match
may result in minor or moderate injury to the user or patient or
the polarities indicated on the battery
damage to the equipment or other property.
compartment.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express
4.
Put the battery cover back in place.
their consent.
Slide the battery cover as indicated,
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
until it clicks into place.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result
in incorrect measurement results.
Note: The measurement
• Make sure that the air tube is not wrapped around other parts of your body
when taking measurements at night. This could result in injury when the air
values continue to be
pressure in the air tube is increased.
stored in memory even
• Do not leave the cuff wrapped on the arm if taking measurements during the
after the batteries are replaced.
night. This could result in injury.
• Do not inflate the arm cuff over 299 mmHg.
2.2 Battery Life & Replacement
• Do not use a mobile phone or other devices that emit electromagnetic fields,
near the unit. This may result in incorrect operation of the unit.
If the battery low symbol (
) appears on the display,
• Do not operate the unit in a moving vehicle (car, airplane).
replace all four batteries at the same time.
• To inflate the cuff manually, refer to Chapter 3.3. If the cuff is over inflated, it
- When the battery low symbol (
can cause internal bleeding.
will still be able to use the unit for a short while. You
should replace the batteries with new ones as soon as possible.
(AC Adapter (Optional) Usage)
- When the symbol (
) remains lit, the batteries are exhausted. You should
• Use only the original AC adapter (optional) designed for this unit. Use of
unsupported adapters may damage and/or may be hazardous to the unit.
replace the batteries with new ones at once. Turn the unit off before replacing
• Plug the AC adapter into the appropriate voltage outlet.
the batteries.
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off
• Remove the batteries if the unit will not be used for three months or more.
the power and unplug the power cord immediately.
• Dispose of batteries according to applicable local regulations.
(Battery Usage)
Four new "AAA" alkaline batteries will last for approximately 300 measurements,
• If battery fluid should get on your skin or clothing, immediately rinse with
when used to take two measurements a day.
plenty of clean water.
• Use only four "AAA" alkaline batteries with this unit. Do not use other types of
Since the supplied batteries are for monitoring use only, they may have
batteries.
a shorter life and do not last for 300 measurements.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace all four batteries at
the same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not take measurements after bathing, drinking alcohol, smoking,
exercising or eating.
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull on the air plug at the connection with the
main unit not the tube itself.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not wash the arm cuff or immerse it in water.
• Read and follow the "Important information regarding Electro Magnetic
Compatibility (EMC)" in the EMC information provided with this unit.
• Read and follow the "Correct Disposal of This Product" in the Technical Data
Section when disposing of the device and any used accessories or optional
parts.
3. Using the Unit
3.1 Applying the Arm Cuff
E
Be sure to wrap the arm cuff correctly so that you get accurate results.
Measurements can be taken in light clothing. However, please remove
thick clothes, such as sweaters, before taking a reading.
Note: You can take a measurement on either your left or right arm. The blood
D
pressure can differ between the right arm and the left arm and therefore
also the measured blood pressure values can be different. Omron
F
recommends to always use the same arm for measurement. If the
values between the two arms differ substantially, please check with your
doctor which arm to use for your measurement.
D. AC adapter jack
(for optional AC adapter)
1.
Insert the air plug into the air jack
E. Battery compartment
on the left side of the main unit.
F. Air jack
2.
Apply the arm cuff to your upper arm. The air tube should be
K
L
centred on the inside of your arm and point down the inside of
M
the arm, so that the air tube runs down the inside of your
N
forearm and is in line with your middle finger.
L.
Pulse display
M.
Deflation symbol
1 to 2 cm
N. Battery low symbol
Note: If your systolic or diastolic
pressure is outside the
standard range
3.
(above 135/85 mmHg),
When the cuff is positioned correctly, close the fabric fastener
the Heartbeat symbol (
)
firmly.
will blink. Refer to
Taking a Reading on the Right Arm
"3.2 Taking a Reading".
Apply the cuff so that the air tube is
If air starts to leak from the arm cuff,
at the side of your elbow.
please replace with a new one.
Refer to "6. Optional Parts".
Notes:
• Be careful not to rest your arm on
O. Arm cuff
the air tube, or otherwise restrict
(Medium cuff:
the flow of air to the cuff.
arm circumference 22-32 cm)
• The cuff should be 1 to 2 cm above
P. Air tube
the elbow.
Q. Air plug
3.2 Taking a Reading
1.
Sit on a chair with your feet flat on
the floor and place your arm on a
table so that the arm cuff will be
at the same level as your heart.
Keep still and do not talk during measurement.
2.
Press the START/STOP button to turn the unit on.
Note: To cancel a measurement, press the START/STOP button to turn the
unit off and to release the air in the arm cuff.
3.
Check the measurement results.
Systolic blood
The unit automatically stores blood
pressure
pressure and pulse rate into its memory.
Diastolic blood
Refer to "3.4 Using the Memory
pressure
Function".
Pulse display
Notes:
• Self-diagnosis of measured results and treatment are dangerous.
Please follow the instructions of your doctor.
• Wait 2-3 minutes before taking another blood pressure reading.
Waiting between readings allows the arteries to return to the condition prior
to taking the blood pressure measurement.
Important:
• If your systolic or diastolic pressure is outside
) starts to blink, you
the standard range, the heartbeat symbol will
blink when the measurement result is displayed.
Recent research suggests that the following
values can be used as a guide to high blood
pressure for measurements taken at home.
Systolic Blood Pressure
Above 135 mmHg
Diastolic Blood Pressure
Above 85 mmHg
This criteria is for home blood pressure measurement.
• Your blood pressure monitor includes an irregular
heartbeat feature. Irregular heartbeats can
influence the results of the measurement.
The irregular heartbeat algorithm
automatically determines if the measurement
is usable or needs to be repeated. If the
measurement results are affected by irregular heartbeats but
the result is valid, the result is shown together with the irregular
heartbeat symbol ( ). If the irregular heartbeats cause the
measurement to be invalid, no result is shown. If the Irregular
heartbeat symbol ( ) is shown after you have taken a
measurement, repeat the measurement. If the Irregular
heartbeat symbol ( ) is shown frequently, please make your
doctor aware of it.
4.
Remove the arm cuff.
5.
Press the START/STOP button to turn the unit off.
Note: If you forget to turn the unit off, it will automatically shut itself off after
five minutes.
4. Handling Errors and Problems
What is Irregular Heartbeat?
Irregular Heartbeat
An irregular heartbeat is a heartbeat rhythm
4.1 Error Messages
that varies by more than 25% from the
Pulse
average heartbeat rhythm detected
Error Display
Blood
while the unit is measuring the systolic
pressure
Irregular Heartbeat
and diastolic blood pressure.
If such an irregular rhythm is detected
Short
Long
more than twice during measurement,
Pulse
the irregular heartbeat symbol (
)
Blood
appears on the symbol when the
pressure
measurement results are displayed.
What is Arrhythmia?
A heartbeat is stimulated by electrical signals that cause the heart to contract.
Arrhythmia is a condition where the heartbeat rhythm is abnormal due to flaws in
the bio-electrical system that drives the heartbeat. Typical symptoms are
skipped heartbeats, premature contraction, an abnormally rapid (tachycardia) or
slow (bradycardia) pulse. This can be caused by heart disease, aging, physical
predisposition, stress, lack of sleep, fatigue etc. Arrhythmia can only be
diagnosed by a doctor through a special examination.
Whether the appearance of the irregular heartbeat symbol (
) in the results
indicates arrhythmia or not can only be determined by an examination and
diagnosis by your doctor.
Warning:
If the irregular heartbeat symbol ( ) is shown frequently, please make your doctor
aware of it. Conducting self-diagnosis and treatment based on measurement
results is dangerous. Be sure to follow the instructions of your doctor.
3.3 Instructions for Special Conditions
If your systolic pressure is known to be more than 220 mmHg, press
and hold the START/STOP button until the arm cuff inflates 30 to
40 mmHg higher than your suspected systolic pressure.
1.
Press the START/STOP button
to turn the unit on.
Measurement starts.
1 to 2 cm
Blinks or appears
continuously
2.
When the cuff starts to inflate, press the START/STOP button
and keep it pressed until the pressure is 30 to 40 mmHg
higher than your suspected systolic pressure.
Note: You cannot inflate the cuff above 299 mmHg. (An error will be displayed
if you try to inflate the cuff above 300 mmHg.)
3.
Release the START/STOP button when the cuff has been
Note: The irregular heartbeat symbol (
inflated to the desired pressure.
The cuff starts to deflate and measurement starts.

4.2 Troubleshooting

4.
The rest of the procedure is the same as for normal
measurement. Refer to "3.2 Taking a Reading", steps 3 to 5.
Note: Do not apply more pressure than necessary.
3.4 Using the Memory Function
This unit has a memory capable of storing 30 sets of readings. Every
The reading is
time you complete the measurement, the unit automatically stores the
extremely low (or
blood pressure and pulse rate.
high).
Note: When 30 sets of readings are stored in memory, the oldest set will be
deleted to store a new set.
1.
Press the Memory (M) button.
The result from the most recent measurement is displayed.
Arm cuff pressure
does not rise.
Important:
If your systolic or diastolic pressure is
Arm cuff deflates too
soon.
outside the standard range, the heartbeat
symbol will blink when the measurement
result is displayed.
Refer to "3.2 Taking a Reading".
Cannot measure or
readings are too low
Note:
If there are no measurements results
or too high.
stored in memory, the screen to the right
is displayed.
The unit loses power
during measurement.
2.
Press the Memory (M) button repeatedly to
cycle through the previous measurement
results.
Nothing happens
when you press the
The Memory number appears for a second
buttons.
before the pulse rate is displayed.
The newest set is numbered "1".
At the end of the measurement, if an irregular heartbeat is
detected, the irregular heartbeat symbol will be displayed
Other problems.
with the result.
3.
Press the START/STOP button to turn the unit off.
If you forget to turn the unit off, it will automatically shut
itself off after five minutes.
To Delete All the Values Stored in Memory
You cannot delete individual stored readings, all the
readings in the unit will be deleted.
1.
To delete stored readings, first press the Memory (M) button.
Then while holding it down, press the START/STOP
button simultaneously for about 2-3 seconds. All readings
will then be deleted.
1
2
2.
Press the START/STOP button to turn the unit off.
If you forget to turn the unit off, it will automatically shut
itself off after five minutes.
5. Maintenance and Storage
Maintenance
To protect your unit from damage, please observe the following:
Cause
Remedy
• Do not subject the main unit and cuff to extreme
temperatures, humidity, moisture or direct sunlight.
Carefully read and repeat the
• Do not fold the cuff or tubing tightly.
Cuff is under inflated.
steps listed under section 3.3.
• Do not inflate the arm cuff over 299 mmHg.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or vibrations
Repeat measurement. Remain
Movement during
still and do not talk during
(for example, dropping the unit on the floor).
measurement.
measurement.
• Do not use volatile liquids to clean the main unit.
Insert the air plug securely.
• Do not wash the arm cuff or immerse it in water.
Air plug disconnected.
Refer to section 3.1.
• Do not use petrol, thinners or similar solvents to clean
the arm cuff.
Apply the arm cuff correctly.
Arm cuff not applied
correctly.
Refer to section 3.1.
• Do not carry out repairs of any kind by yourself.
If a defect occurs, consult your local OMRON
Clothing is interfering
Remove any clothing
representative.
with the arm cuff.
interfering with the arm cuff.
Air is leaking from the
Replace cuff with new one.
arm cuff.
Arm cuff has been
Do not inflate the arm cuff
inflated above
above 299 mmHg.
• The unit should be cleaned with a soft, dry cloth.
299 mmHg.
Refer to section 3.3.
• Use a soft, moistened cloth and soap to clean the arm cuff.
Note: Read and follow the "Correct Disposal of This Product" in the
Technical Data Section when disposing of the device and any
used accessories or optional parts.
Replace all four "AAA"
Battery power is low.
batteries with new ones.
Calibration and Service
Refer to section 2.1.
• The accuracy of this blood pressure monitor has been
carefully tested and is designed for a long service life.
• It is generally recommended to have the unit inspected every
two years to ensure correct functioning and accuracy.
Contact your local OMRON
Please consult your local OMRON representative.
Device error.
representative.
Storage
) may also be displayed with error
1.
messages.
Unplug the air tube from the air jack.
Gently fold the air tube
into the arm cuff.
Note: Do not bend the air tube excessively.
Problem
Cause
Remedy
Do not store the unit in the following situations:
Apply the arm cuff
correctly.
• If the unit is wet.
Arm cuff not applied
correctly.
Refer to section 3.1.
• Locations exposed to extreme temperatures, humidity, direct sunlight,
dust or corrosive vapours.
Remain still and do not
Movement or talking
• Locations exposed to vibrations, shocks or where it will be at risk of
talk during
during measurement.
falling.
measurement.
Remove any clothing
6. Optional Parts
Clothing is interfering
interfering with the arm
with the arm cuff.
cuff.
Arm Cuff
Arm circumference
Arm circumference
Make sure that the air
The air plug is not
22 - 32 cm
32 - 42 cm
plug is connected
securely inserted
securely.
into the main unit.
Refer to section 3.1.
Air is leaking from the
Replace the arm cuff
arm cuff.
with a new one.
Medium Arm Cuff
Large Arm Cuff
Apply the cuff correctly
Model: HEM-CR24
Model: HEM-CL24
so that it is firmly
wrapped around the
The arm cuff is loose.
Using the Optional AC Adapter
arm.
Refer to section 3.1.
1.
Insert the AC adapter plug
Inflate the cuff so that it
into the AC adapter jack on
The arm cuff has not
is 30 to 40 mmHg
above your previous
been inflated
the side of the main
measurement result.
sufficiently.
unit.
Refer to section 3.3.
2.
Plug the AC adapter into an
The batteries are
Replace the batteries
empty.
with new ones.
electrical outlet.
The batteries are
Replace the batteries
To disconnect the AC adapter, unplug the AC adapter from the
empty.
with new ones.
electrical outlet first and then remove the AC adapter plug from the
main unit.
The batteries have
Insert the batteries with
been inserted
the correct (+/ -)
incorrectly.
polarity.
Press the START/STOP button and repeat
measurement.
If the problem continues, try replacing the
batteries with new ones.
If this still does not solve the problem, contact
your local OMRON representative.
7. Technical Data
Product Description
Automatic Blood Pressure Monitor
Model
JPN2
Display
LCD Digital Display
Measurement Method
Oscillometric method
Measurement Range
Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats/min.
Memory
30 Measurements
Accuracy
Pressure: ±3 mmHg
Pulse: ±5% of reading
Inflation
Fuzzy-logic controlled by electric pump
Deflation
Automatic pressure release valve
Rating
DC6V 4W
Power Source
4 "AAA" batteries 1.5V
or AC adapter (optional, INPUT AC100-240V
50/60Hz 0.12A)
Battery Life
Approx. 300 measurements
(using new alkaline batteries)
Applied Part
= Type B
Protection against
Internally powered ME equipment
Electric Shock
(When using only the batteries)
= Class II ME equipment
(Optional AC adapter)
Operating
10°C to 40°C/ 30 to 85% RH
Temperature/Humidity
Storage Temperature/
-20°C to 60°C/ 10 to 95% RH/ 700-1060 hPa
Humidity/ Air Pressure
Console Weight
Approx. 250g without batteries
Cuff Weight
Approx. 120g
Outer Dimensions
Approx. 104 (w) mm
×
84 (h) mm
×
129 (l) mm
Cuff Dimensions
Approx. 146 mm
×
446 mm
(Medium cuff: arm circumference 22 to 32 cm)
Cuff Material
Polyester and nylon
Package Content
Main unit, medium cuff, instruction manual, EMC
information, battery set
Note: Subject to technical modification without prior notice.
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
• This blood pressure monitor is designed according to the European Standard
EN1060, Non-invasive sphygmomanometers Part 1: General Requirements
and Part 3: Supplementary requirements for electromechanical blood
pressure measuring systems.
• This OMRON product is produced under the strict quality system of OMRON
HEALTHCARE Co., Ltd., Japan. The core component for OMRON blood
pressure monitors, which is the Pressure Sensor, is produced in Japan.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it
should not be disposed of, with other household wastes at the end
of its working life. To prevent possible harm to the environment or
human health from uncontrolled waste disposal, please separate
this product from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
AC Adapter
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they can
return this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions
of the purchase contract. This product should not be mixed with other commercial
wastes for disposal.
Adapter S
Model: 60240HW5SW
Manufacturer
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto
617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp
THE NETHERLANDS
Asia Pacific HQ
OMRON HEALTHCARE SINGAPORE PTE LTD.
438A Alexandra Road, #05-05/08, Alexandra Technopark
Singapore 119967
www.omronhealthcare-ap.com
Production Facility
OMRON HEALTHCARE Co., Ltd.
Matsusaka Factory
Mie, JAPAN
Made in Japan