DreamStation BiPAP autoSV User Manual
Alert
Pair?
Yes/No
Flow Sensor Error:
Limited Therapy
SD Card Removed.
Oximetry: Good
Study
(icon only)
SD Card Error:
Remove and
Reinsert
38
Alarms, alerts, and troubleshooting
Icon
Priority
Status
Notification
Notification
or Alert
Notification
Notification
Description
Prompts to accept or
decline pairing to a
Bluetooth compatible
device.
Flow sensor malfunction.
Indicates SD card has been
removed from therapy
device and not reinserted
before the start of the
current therapy session.
After exiting therapy, the
therapy device will display
the Good Oximetry pop-
up screen if the oximetry
sensor maintained a good
connection for at least 4
consecutive hours within a
24 hour window with breaks
in therapy no greater than 1
hour each.
SD card error detected.
Device cannot read the SD
card.
A problem may exist with
the SD card or it was
ejected during a writing
activity, or it was inserted
incorrectly.
User action
Rotate the control dial to
accept pairing (Yes), or
decline (No), then press
the control dial to confirm
selection. Selecting Yes
opens the Bluetooth
passkey screen.
Press the control dial to
acknowledge and clear
the message. Disconnect
the device from power.
Immediately remove
patient from ventilator and
connect them to alternate
source of ventilation.
Contact your home care
service provider for service.
Reinsert SD card, or press
the control dial to clear
alert.
Press the control dial to
acknowledge and clear the
message.
Remove SD card and
reinsert. If alert continues
to occur, replace with
another card or contact
your home care service
provider.