Programming examples: long integration measurements
The following command sequence will trigger and return one long integration measurement:
Battery channel (#1)
Command
SENS:RANG 5
VOLT 15
CURR 0.75
OUTP ON
SENS:LINT:TLEV:RANG 0.5
SENS:LINT:TEDG RISING
SENS:LINT:TLEV:ONE 0.1
SENS:LINT:TIME:AUTO
SENS:FUNC "LINT"
READ?
Charger channel (#2)
Command
SENS2:RANG 5
SOUR2:VOLT 15
SOUR2:CURR 0.75
OUTP2 ON
SENS2:LINT:TEDG RISING
SENS2:LINT:TLEV 0.1
SENS2:LINT:TIME:AUTO
SENS2:FUNC "LINT"
READ2?
Description
' Select 5A range.
' Set output voltage to 15V.
' Set current limit to 750mA.
' Turn output on.
' Select 1A trigger level range.
' Select rising trigger edge to initiate measurement.
' Set trigger level to 100mA for 1A trigger level range.
' Set integration time automatically for single pulse.
' Select long integration function.
' Trigger and return one reading and reading shown on
display.
Description
' Select 5A range.
' Set output voltage to 15V.
' Set current limit to 750mA.
' Turn output on.
' Select rising trigger edge to initiate measurement.
' Set trigger level to 100mA.
' Set integration time automatically for single pulse.
' Select long integration function.
' Trigger and return one reading and reading shown on
display.
Front panel operation