12. Hazardous Area considerations

12.1 Ex Safety aspects and IP Protection (Europe)

According to ATEX Directive (European Directive 94/9/EC of 23
March 1994) and relative European Standards which can
assure compliance with Essential Safety Requirements, i.e.,
EN 60079-0 (General requirements) EN 60079-1 (Flameproof
enclosures "d") EN 60079-11 (Equipment protection by intrinsic
safety "i") EN 60079-26 (Equipment with equipment protection
level -EPL- Ga) EN 61241-0 (General requirements)
EN 61241-1 (Protection by enclosures "tD") EN 61241-11
(Protection by intrinsic safety"iD") the pressure transmitters of
the 2600T SERIES have been certified for the following group,
categories, media of dangerous atmosphere, temperature
classes, types of protection. Examples of application are also
shown below by simple sketches.
a) Certificate ATEX II 1 G Ex ia IIC T4/T5/T6 and II 1 D Ex iaD
20 T85°C
FM Approvals certificate number
FM09ATEX0024X (Lenno products)
FM09ATEX0069X (Minden products)
FM11ATEX0035X (Bangalore products)
FM11ATEX0046X (Shanghai products)
The meaning of ATEX code is as follows:
— II : Group for surface areas (not mines)
— 1 : Category
— G : Gas (dangerous media)
— D: Dust (dangerous media)
— T85°C: Maximum surface temperature of the transmitter
enclosure with a Ta (ambient temperature) +40°C for Dust
(not Gas) with a dust layer up to 50 mm depth.
Application for pressure transmitter Ex ia categories 1G and 1D
Application with Gas
Zone 0
266 Tx Category 1G Ex ia
Note: the transmitter must be
connected to a supply
(associated apparatus)
certified [Ex ia]
12 Hazardous area considerations
Certificate IECEx Ex ia IIC T4/T5/T6 and Ex iaD 20 T85°C
FISCO, IP67
IECEx certificate number
IECEx FME 09.0003X (Lenno and Minden products)
IECEx FMG 11.0019X (Shanghai products)
Important. The number close to the CE marking of the transmitter
safety label identifies the Notified Body which has responsibility for
the surveillance of the production.
The other marking refers to the protection type used according
to relevant EN standards:
— Ex ia: Intrinsic safety, protection level "a"
— IIC: Gas group
— T4: Temperature class of the transmitter
(corresponding to 135°C max) with a Ta from -50°C to +85°C
— T5: Temperature class of the transmitter
(corresponding to 100°C max) with a Ta from -50°C to +40°C
— T6: Temperature class of the transmitter
(corresponding to 85°C max) with a Ta from -50°C to +40°C
About the applications, this transmitter can be used in "Zone 0"
(Gas) and "Zone 20" (Dust) classified areas (continuous hazard)
as it is shown on the following sketches.
Application with Dust
2600T Series Pressure transmitters | OI/266/FF-EN Rev. B 47
Zone 20
266 Tx Category 1D IP6x (Ex ia)
Note: the protection is mainly
assured by the "IP" degree
associated to the low power
from supply. This can either be
[ia] or [ib] certified [Ex ia]