approximately 7 seconds). At this point release the button and the instrument is
ready for a new measurement.. The displayed value is the last recorded power
measurement carried out at this wavelength setting.
carry out a measurement of a known laser source. Shift the switch on + or – if
you need to increase or decrease the reading displayed on the LCD to match your
reference. This is done by limited pushes on the ON/Zero key, until the desired
power value is achieved on the display. Reset the switch to its middle position
CAL and re-cap. The instrument is ready for new measurements.
6
9-Out of Range Notice
When the head is exposed to a source exceeding its
full scale, the LCD shows four short lines instead of a
numeric value.
10-Wavelength Selection
To select the wavelength of operation (e.g. from CO2 to
YAG), go back to Paragraph 2 (shut the instrument off),
but keep the ON/Zero button pushed until the wavelength
change appears on the display (this step takes
11-Customer's calibration
This instrument can be re-calibrated by the user to
match a custom reference. Remove the plastic cover of
the adjustment switch located at the right side of the
instrument body. Zero the instrument as indicated in
Paragraph 2 of this Operation Section and afterwards