5. Error Codes
These are displayed in the LCD display as E ##.
The free-piston Stirling engine is not maintaining a working temperature. This
could arise from blocked filters, blocked air passages, unusually hot ambient
10
temperatures (for example >45 º C), or a fan failure. Check air passages and clear
any blockages for good air circulation. Remove from hot environments (car
interiors, etc.).
A component which monitors free-piston Stirling engine performance has failed.
40
The free-piston Stirling engine cannot regulate itself. Please return for repair.
6. Maintenance
6.3 – Storage
1. Turn off the power and allow the freezer to come to room temperature.
2. Dry the inside of the freezer compartment and clean any spills.
3. Disinfect with suitable sterilizing agent if the freezer has been used for biohazards.
6.4 – Cleaning
The exterior surfaces of the freezer can be cleaned as needed by using a soft
cloth and mild detergent. Do not use solvent or harsh abrasive cleansers or pads.
18 of 28
Operating Manual:
6.1 – Caring for Your
Freezer Gasket
The temperature differential between
ambient and the ultra-low interior can attract
moisture near the gasket. For best results
always wipe away moisture during openings
to prevent ice accumulation. Do not use sharp
instruments to scrape ice from the surface.
6.2 – Air Intake Filter
Clean accumulated dust and dirt on the
air intake filter every month. Pull the filter
cover to the left as shown. Please note that it
cannot be completely removed. Gently clean
the filter with a vacuum cleaner. If there are
stubborn residues use a soft brush to work
them loose.
ULT25 NE
10327.1 | 12.22.2016
Need help?
Do you have a question about the Shuttle ULT25NE and is the answer not in the manual?