You can now move on to either Auto Calibration or Table Calibration.

2 Auto Calibration

The automatic calibration requires that you apply two known (ascending) weights in sequence. Just enter the
required engineering unit value for the two weights in the fields marked 1 and 2.
Next, apply the lowest weight and allow a settle time. Click the 'Acquire' button then apply the higher weight. After
another settle time click the second 'Acquire' button and the calibration is complete.
3 Table Calibration
This option lets you calibrate direct from the loadcell manufacturers certificate without having to apply weights to
the input.
Choose 2 points from the loadcell certificate and enter the mV/V values and the corresponding weights on the
certificate (or convert the certificate weight to any other engineering unit of your choice) click the 'Calibrate' button
and the calibration is complete.
Mantracourt Electronics Limited - DSCUSB Advanced Manual