Linearisation - Mantracourt DSCUSB Advanced User's Manual

Digital sensor card usb – strain gauge input
Hide thumbs Also See for DSCUSB:
Table of Contents

Advertisement

4 Linearisation

This stage is optional. If you do not want to use linearization then select Disabled from the drop down list that is
presented when you click the Number of linearization points field.
To use linearization select the number of points you wish to use from 2 to 7. Once selected this will enable the
appropriate number of text fields for you to enter the known weights you are going to apply. You should enter these
in the same engineering units that were used in either the Auto Calibration or the Table Calibration.
The linearization points are usually spread out evenly between the two points used for auto calibration but if you
know that a particular section of the range is less linear you may want to concentrate the points in that area.
Once you have entered the engineering units weights for each point you can apply the weight, wait a settle time
then click the appropriate Acquire button.
Once you have completed all points the following message will be displayed.
9
Mantracourt Electronics Limited - DSCUSB Advanced Manual