Mantracourt DSJ4 Instruction Sheet

Strain gauge or load cell digitiser module 4 channel

Advertisement

Quick Links

Instruction Sheet
Features
LED Power Indicator
Option to terminate bus through 120 ohm resistor
Bus Connections for communication and Power Supply
Bus Connection via 2 Part screw connectors
'Daisy Chain' Bus through D type connectors
2 Part Connectors for Load Cells
CAN Compliant
3 year guarantee
IP65 NEMA 4 ABS surface mounting case (as shown)
Introduction
The function of the DSJ4 is to enable the easy connection for 4 channels of a load cells and DSC Cards for
communications to a PC or PLC.
Designed to take 4 DSC cards in a single enclosure, the DSJ4 offers a convenient and practical solution to the
installation of digital load cells with platforms, silos and any weighing systems where connection to PC and PLC's is
an essential requirement of the system.
Supplied as an OEM device 100 x 185mm PCB, it has options for fitting in an IP65 ABS case, or to a DIN rail fixture.
Notes for installation & preparing for use
For the operational and communication requirements of the
DSC card, refer to the DSC User Manual Instructions.
The DSC card is mounted on the main PCB with 3 pillars.
The D type connectors J10 & J11 and the 5 way two-part
connector J9 provide the power connections and the RS485 or
CAN connections. J10 and J11 also provide a 'daisy chain'
interconnect facility for multiple DSJ4 units.
J5 to J8 are two-part 4 way connectors which provide
connections for the digital input and output. J1 to J4 are 8
way two-part connectors providing connections to the strain
gauge and external temperature sensor.
DOC-DSJ4-1 DSJ4 Instruction Sheet
Strain Gauge or Load Cell Digitiser Module (4 channel)
Issue 1.8 27.05.2015
mantracourt.com
Photo showing DSJ4 with 4 x DSC pcb cards
fitted & in ABS surface mounting case
Connection details are shown on the layout diagram on the
reverse of this sheet.
Cable Screening: It is strongly recommended that all cable
screens are connected.
A green LED L1, marked PWR on the PCB provides an indication
of power.
Link LK1 is used to provide a 120R terminating resistor for
RS485 or CAN connections.
sales@mantracourt.com
tel +44 (0)1395 232020
DSJ4

Advertisement

loading

Summary of Contents for Mantracourt DSJ4

  • Page 1 & in ABS surface mounting case Introduction The function of the DSJ4 is to enable the easy connection for 4 channels of a load cells and DSC Cards for communications to a PC or PLC. Designed to take 4 DSC cards in a single enclosure, the DSJ4 offers a convenient and practical solution to the installation of digital load cells with platforms, silos and any weighing systems where connection to PC and PLC’s is...
  • Page 2: Power Supply Specifications

    When using J10 & J11 for communications the cable screen is connected to pin 1. ISO 9001 REGISTERED FIRM In the interests of continued product development, Mantracourt Electronics Limited reserves the right to alter product specifications without prior notice. DOC-DSJ4-1 DSJ4 Instruction Sheet Issue 1.8 27.05.2015 mantracourt.com...