This page allows you to calibrate the module using various methods, set limits, perform system zero and set units
description.
The page is separated by tabbed sections. Click in a tab to change sections.
Revert to factory default calibration – Click the 'Reset' button to remove any custom calibration you have applied.
If the DSC module supports it the date of the last change to calibration settings will be displayed.
There are two main calibration methods. You can choose Basic Calibration for a simple 2 point auto or table
calibration or choose Compensated Calibration if you are going to perform linearization or temperature
compensation.
Basic Calibration
Choose this instead of Compensated Calibration if you do not want to perform temperature compensation or
linearization.
Auto Calibration
The automatic calibration requires that you apply two known (ascending) weight in sequence. Just enter the required
engineering unit value for the two weights in the fields marked 1 and 2.
Next, apply the low weight and allow a settle time. Click the 'Acquire' button then apply the high weight. After
another settle time click the second 'Acquire' button and the calibration is complete.
Table Calibration
5
Mantracourt Electronics Limited - DSCUSB Advanced Manual