Section 2 Specifications - Thermo Scientific MaxQ 5000 Operating Manual And Parts List

Incubated and refrigerated floor shakers
Table of Contents

Advertisement

Section 2
Specifications
4348/4354
Model No.
Motor
Permanent Magnet DC
Speed
None
Speed Shut Off
None
Timer
None
Unbalanced Load
None
2-2
MaxQ 5000 Shaker
4349/4355
Permanent Magnet DC
None
None
None
None
Optional Platform Dimensions in. (cm)
Catalog Number
30115
Intended Use
Orbital shakers are designed to provide increased aeration in a stable
temperature environment.
Unintended Use
1) Not intended for use in Class I or II applications as defined in 21 CFR
2) Not intended for mixtures of flammable materials
4350/4358
Solid State Brushless DC
Audible with flashing LED
indicates when speed devi-
ates more than 10% of set
point
When speed deviates 10%
of set point, unit will shut
down immediately
Beeps twice when time has
expired. Shaking motion
stops.
If the unit is running in an
unbalanced condition, an
alarm will sound and the
shaker will stop until the end
user corrects the condition.
The speed display will flash
"bAL" on speed panel LED.
L xW
Universal 30" x 18" (76.2 x 45.7 cm)
4351/4358
Solid State Brushless DC
Audible with flashing LED
indicates when speed devi-
ates more than 10% of set
point
When speed deviates 10%
of set point, unit will shut
down immediately
Beeps twice when time has
expired. Shaking motion
stops.
If the unit is running in an
unbalanced condition, an
alarm will sound and the
shaker will stop until the end
user corrects the condition.
The speed display will flash
"bAL" on speed panel LED.
Thermo Scientific