Degassing
For initial cleaning solution degassing.
Step
Action
1
Turn Main Power switch on.
2
Turn HEAT switch on.
3
Turn the Timer Knob clockwise to 5–10 and let the unit run to
allow the solution to "degas."
NOTE: Refer to page 39 for information on degassing.
Cleaning Items (Treating Samples)
NOTE: To stop ultrasonics at any time, turn the Timer Knob to the zero
position.
Step
Action
1
Turn Main Power switch on.
2
Turn the Timer Knob clockwise to set the amount of time
(0–60 minutes) you wish the items to be cleaned.
Turn the Timer Knob counterclockwise to the HOLD position for
continuous operation.
3
Place the items into a basket, perforated tray, or beakers in a
positioning cover.
4
If using beakers or a solid tray, add cleaning solution to
beakers or tray to cover the items.
5
Slowly lower the tray or beakers into the tank. Do not allow
items to contact the tank bottom.
6
When items are clean, slowly remove them from the tank.
7
Rinse the clean items with clean water and dry them, if
necessary.
MH Series
Operating your Unit
16 Operating Your Ultrasonic Bath