The Li-Ion Battery Pack

The factory-supplied Li-ion battery pack is designed to last for 16 hours
of normal operation between charges (without alarm or back light
conditions). The rechargeable batteries have a 1-year warranty. Age,
ambient temperature, and heavy usage may impact battery life.
Battery packs slowly drain even if the monitor is turned off. Therefore,
it is strongly recommended that the EntryRAE be stored on its charging
cradle. The battery can trickle-charge when not in use.
If the battery packs have not been charged for 10 days, the battery
voltage will be low. Fully charge battery packs before going into the fi eld,
and recharge them upon returning from the fi eld.
To reduce the risk of ignition of hazardous atmospheres, recharge
!
battery only in areas known to be non-hazardous. Remove and replace
battery only in areas known to be non-hazardous.
Ne charger les batteries que dans emplacement désignes non
dangereux.

Programming Mode

In addition to calibration, authorized users may change the monitor
settings to their requirements using the Programming Mode.
Note: Monitoring gas concentrations pauses during Programming Mode
and during calibration. Datalogging also pauses during Programming
Mode, but resumes when programming is fi nished.

Security

EntryRAE provides the option of securely enabling a password to
access Programming Mode. The default password is "0000." To enable
password access, change the "0000" to a desired password. (Refer to
"Change Pass?" on page 26.)