Aesculap Power Systems
Acculan® 3Ti Dermatome GA670
Press on nut 2 and push flap rod 18 approx. 4 mm
to one side.
Twist flap rod 18 until it can be removed.
Remove flap rod 18.
Slide off flaps 12 from flap rod 18.

Connecting the accessories

Risk of injury due to unapproved
configuration
components!
DANGER
For all components, ensure that
their classification (e.g. Type BF
or Type CF) is the same as that
of the device in the respective
application.
Combinations of accessories that are not mentioned in
the present instructions for use may only be employed
if they are specifically intended for the respective
application, and if they do not compromise the
performance and safety characteristics of the
products.
Also, all equipment connected at the interfaces must
demonstrably meet the respective IEC standards (e.g.
IEC 60950 for data processing equipment, EN 60601
for electromedical devices).
All configurations must meet the system standard
IEC 60601-1-1. The person connecting the units is
responsible for the configuration and must ensure
compliance with system standard IEC 60601-1-1 or
equivalent national standards.
Please contact your B. Braun/Aesculap Partner or
Aesculap Technical Service, address see Technical
Service with any inquiries in this respect.
6

Inserting the dermatome blade

using
additional
Risk of infection and contamination
due to torn or cut surgical gloves!
Avoid contact with the cutting
edge
when
DANGER
dermatome blade.
Risk of injury and material damage
due to inadvertent activation of the
dermatome while inserting the
dermatome blade!
DANGER
Prior to inserting the dermatome
blade, engage the safety catch.
Engage safety catch 5.
Push both clamping levers 7 in the direction of the
arrows.
Blade cover 6 is now unlocked.
Open blade cover 6 in the direction of the arrow.
Safely insert dermatome blade 24 in pusher dog 10
and blade guide peg 9, see Fig. 2.
Close blade cover 6.
Tighten both clamping levers 7 so that roll pin 17
engages in the recess of clamping lever 7.
inserting
the