Intraoperative Battery Change - Braun Acculan 3Ti Instructions For Use/Technical Description

Veterinary surgical power tool
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Aesculap Power Systems
Acculan® 3Ti Dermatome GA670
Fig. 8
Removing the battery
Note
The battery must be removed after the surgical
operation and prior to processing.
Note
Battery removal device can be used to facilitate the
removal of the battery, see Fig. 9!
Hold dermatome 1 with the battery compartment
facing upward.
Simultaneously press both cover release buttons 22
at cover 21 and remove cover 21.
With one hand, completely envelope the battery
compartment at its lower end, see Fig. 7.
Tap battery 23 against the palm of the other hand
and remove the battery from the battery
compartment, see Fig. 8.
10

Intraoperative battery change

Note
The function checks must be carried out prior to each
surgical application of the unit and after each
intraoperative battery change!
Battery removal device is used for intraoperative
battery changes ensuring that sterile conditions are
safely maintained.
25
Fig. 9
Using the battery removal device
Hold dermatome 1 with the battery compartment
facing upward and remove cover 21.
Mount sterile battery removal device 25, see Fig. 9.
1

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents