Acculan® 3Ti Dermatome GA670

10. Technical specifications

Model
DC voltage
Rated current
Max. power
Oscillation frequency
Blade travel
Cutting depth
Cutting width
Weight (operational)
Dimensions L x W x H
Application part
EMC
Conforming
to
standard
Classification acc. to
directive 93/42/EEC
Model
Cell type
DC voltage
Nominal capacity
Weight
Dimensions L x W x H
EMC
24
GA670
9.6 V
approx. 2 A
250 W
0 to approx. 6 500 1/min
3.1 mm
0.2 mm to 1.2 mm,
1/10 mm dial gradation
max. 78 mm, set by
10 flaps of width 8 mm
approx. 1.7 kg
approx. 290 x 110 x
90 mm
Type BF
IEC/DIN EN 60601-1-2
IEC/DIN EN 60601-1
IIa
GA666
NiMH
9.6 V
1.05 Ah
approx. 0.304 kg
approx. 120 x 43 x
50 mm
IEC/DIN EN 60601-1-2
Model
Conforming to standard
Classification acc. to
directive 93/42/EEC

10.1 Ambient conditions

Storage and transport conditions
Relative humidity
Ambient temperature
Atm. pressure
Operating conditions
Relative humidity
Ambient temperature
Atm. pressure
GA666
IEC/DIN EN 60601-1
I
45 % to 85 %
-20 °C to +50 °C
(up to 30 days)
-20 °C to +40 °C
(up to 90 days)
-20 °C to +30 °C
(up to 1 year)
(no direct sunlight)
50 kPa to 106 kPa
30 % to 75 %
+10 °C to +40 °C
(no condensation)
70 kPa to 106 kPa