Video Test Using Sc600 Scope Calibration Option - Fluke 190-104 Service Manual

Scopemeter 190 series ii
Hide thumbs Also See for 190-104:
Table of Contents

Advertisement

Fluke 190-104/190-204
Service Information
4119-75
75 OHM
8. Select the following Test Tool setup:
9. Using
10. Observe the trace, and check to see if the Test Tool triggers on line number 310 (PAL
Figure 4-17. Trace for PAL/SECAM line 310

4.6.18 Video test using SC600 Scope Calibration Option

You can skip this test if you did test
Generator.
Only one of the systems NTSC, PAL, PALplus, or SECAM has to be verified.
Proceed as follows:
4-30
PM9075
A
B
C
D
Figure 4-16. Test Tool Input A to TV Signal Generator Inverted
F4
Press
to open the Trigger Options menu.
Choose
VIDEO on A...
Polarity: NEGATIVE | PAL
select line number 310 (PAL or SECAM) or 262 (NTSC)
or SECAM, see Figure 4-17), or line number 262 (NTSC, see Figure 4-18).
Negative Video
USE 75 OHM TERMINATION
USE 75 OHM COAX CABLE
PM5418
, then from the shown opened menu choose
( or
or
NTSC
PALplus
Figure 4-18. Trace for NTSC line 262 Negative
4.6.17 Video test using the Video Pattern
PM9082
PM9081
or
)
SECAM
Video