Specifications
Single Channel Amplitude
Single Channel Amplitude
Absolute amplitude accuracy (FFT)
(A combination of full scale accuracy, full
scale flatness, and amplitude linearity.)
FFT full scale accuracy at 1 kHz (0 dBfs)
FFT full scale flatness (0 dBfs) relative to 1
kHz
FFT amplitude linearity at 1 kHz
Measured on +27 dBVrms range with time
average, 0 to -80 dBfs.
Amplitude resolution (16 bits less 2 dB
over-range) with averaging
Residual dc response
FFT mode frequency display
(excludes A-weight filter)
1-4
±2.92% (0.25 dB) of reading
±0.025% of full scale
±0.15 dB (1.74%)
±0.2 dB (2.33%)
±0.58% (0.05 dB) of reading
±0.025% of full scale
0.0019% of full scale (typical)
<-30 dBfs or -66 dBvdc (.5mVdc)(whichever is greater)
Agilent 35670A