Agilent 35670A
To set up the harmonic distortion test
Performance Test only
This test verifies that the Agilent 35670A meets its FFT dynamic range
specification for harmonic distortion. In this test, a low distortion oscillator or a
frequency synthesizer and 24.5kHz notch filter outputs a signal to all channels. The
second, third, fourth, or fifth harmonic is then measured. If the harmonic falls
outside the analyzer's frequency range, the analyzer measures the alias frequencies.
The test records at the end of this chapter list the fundamental frequencies. If you
are using the synthesizer and notch filter, the frequencies listed in the test record are
approximate.
1
2 ch
Using an HP 339A
1
4 ch
Using an HP 339A
Verifying Specifications
To set up the harmonic distortion test
3-25