OPERATION
6
Level
Designation
(available if Maes.Inst =
Mass, see menu item
1.1.3)
3.3
3.3.1
Poll. Adr.
3.3.2
HART SV
3.3.3
HART TV
3.3.4
HART 4V
3.4
3.4.1
Fluid 1
3.4.2
Medium
(available if Fluid = Gas or
Liquid, see menu item
3.4.1)
(available if Fluid = Steam,
see menu item 3.4.1)
1 Density check
3.5
3.5.1
T-sensor
3.5.3
Sat.P/T
(verfügbar wenn FLUID =
DAMPF, siehe Menüpunkt
3.4.1)
3.5.4
Temp. Opr.
52
Selection / entry
Explanation
kg
Select totalizer unit for mass flow measurement
Unit ↑...^
0000000000
Enter preset totalizer value
kg ↑...↑...↑...^
Reset yes /
Reset totalizer /
Reset no ↑...^
Do not reset totalizer/
Disp. off /
Display totalizer /
Disp. on ↑...^
Do not display totalizer
HART settings
000
Enter HART polling address for "Multidrop operation"
0 – 15
↑...→↑...→↑...^
Total Flow ^
HART secondary variable
Temp. ^
HART tertiary variable
Density ^
Density ^
HART quaternary variable
Temp. ^
Set fluid and medium
Set fluid type
Gas ↑...
Gas
Liquid. ↑...
Liquid
Steam ↑...^
Steam
Set medium
Custom ↑...
Customer medium
Sat.Steam ↑...^
Saturated steam
Custom
Set pressure, temperature and density
Internal temperature sensor
No ↑...
No temperature sensor available
Yes ↑...^
Temperature sensor available
Density calculation for saturated steam
Sat. T ^
Density calculation for saturated steam with the
saturated temperature
Operating temperature
°C
Set temperature unit
Unit ↑...^
0000000.0
Operating temperature
°C ^
Disp. on /
Display temperature /
Disp. off ↑...^
Do not display temperature
www.siemens.com/flow
SITRANS FX300
06/2009 - SFIDK.PS.050.F2.02 A5E02100423