5. Transceiver menus
5.3.3 Resistive leakage current measurement
Not available in all versions.
Select sensor ID to read data from with the
key.
Start the reading by pressing
Before the transmission begins the operating voltage must be
given. (see 5.3.3.1)
Progress bar
Error code
A successful transmission is marked with
A failed transmission is marked with
If the transmission was unsuccessful, an error code
will be displayed.
Back to read data menu
5.3.3.1 Operating voltage
Not available in all versions.
The actual operating voltage must be given when
measuring the resistive leakage current.
The system voltage is displayed as a default value.
Adjust the value to the actual value with
and
keys.
Enter the correct value by pressing
NOTE! If the system is 3-phase, the value should be the actual
phase-phase voltage. Otherwise, for 1-phase systems, the value
should be the phase-ground voltage.
The progress bar will indicate that the transmission
has begun.
47
1HSA 801 080-15en EXCOUNT-II Users Manual