Probe Length 1.0m to 2.0m
Step
Action
Fully thread the nut onto the central conductor of the process
connector. Place a lock washer between the locknut and the rod
segment and torque the nut against the rod segment and lock
1
washer to secure the connection.
Note: Tighten rod connection point to the following torque:
Slide a spacer onto first rod. Connect the rod "end" segment to first
rod using a stud and lock washer. Refer to Detail A of
2
Note: Tighten rod connection point to the following torque:
Slip the coaxial outer tube over the rod and tighten to the process
connector.
Note: Tighten connection point to the following torque:
3
It is recommended that a process compatible thread locking
compound (i.e. Loctite 242) be used on the outer conductor
threaded joints.
Insert an end spacer into the end of tube. Align the holes in the
end spacer with the holes in the outer tube and insert the 2 locking
4
pins. Refer to Detail B of
Figure 5: SLG720 Coaxial Probe
Assembly
December 2017
SLG720 6.0 Nm (4.4 ft-lbs)
SLG720 6.0 Nm (4.4 ft-bls)
SLG720 30 Nm (22 ft-lbs)
Figure
.
Quick Start Installation Guide
5.
Figure 5.