Specifications; Control Outputs; Operating Conditions - Process Technology DQ15D Instruction Manual

Digital temperature control
Table of Contents

Advertisement

DQ15D Specifications
Standard Input
Input Range
Set Point Range
Sensor Break or
Short Protection
Accuracy
Enclosure
Display
Control Function

Control Outputs

ON/OFF Differential
Memory
Supply Voltage

Operating Conditions

Input Options
Output Options
M-34-01-06 3/31/2016
Equipment protected throughout by Double Insulation or Reinforced Insulation.
2 wire1000 ohm RTD TCR (alpha),
0.00385 ohm/ohm/°C
RTD Self Heating Coefficient:
5° C/w in 0.2 m/s water; 200° C/w in
1 m/s air measurement current, 0.1 to 0.2 mA
-40°F (-40°C) to 1000°F (538°C); °F or °C field selectable
Selectable throughout the input range
De-energize control output (No sensor short
protection with Thermocouple sensor)
± 0.25% span, ± 1 digit
Type 12, IP54
Face suitable for panel mounting (#20 ga. through 1/4 thick panels)
4 digit, (1/2" nominal), LED display screen
ON/OFF Electromechanical Relays
SP1 Set Point
(reverse acting) SPDT 20A resistive@240 VAC max
1HP@240 VAC max, 1/2 HP@120VAC
SP2 Set Point
(direct acting) SPDT 2A resistive@240 VAC max
1/10HP@240 VAC max, 1/20 HP@120VAC
Field adjustable, 1° (F or C) to 99°
Nonvolatile
85 to 265 VAC, 50-60 Hz, 13VA
• Indoor Use Only
• 20°F (-7°C) to 140°F (60°C)
• Max Altitude: 3000m
• Max Relative Humidity: 80%
• Pollution Degree 2
• Installation Category II
RTD 2 & 3-wire 100 ohm 0.00385 ohm/ohm/ºC or
0.00392 ohm/ohm/ºC
Thermocouples - (types J, K, T, R)
NIST Monograph 175, revision ITS-90 (Obsolete)
Current 4-20mA DC
Voltage (1-10 VDC)
Frequency (0-200 Hz, counts/second),
+/- 6 VDC p to p (up to 30VDC peak with Sw2 ON)
4-20 mA DC proportional to the process (display) value.
Maximum impedance 450 Ohms.
1
DQ15D/T-DQ15D Manual

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

T-dq15d

Table of Contents