90
Add Strip Lot
Strip Lot Confi rmation
D-530: Do you want to
use the suggested
values for strip 545603?
Use by: 12/10/17
L1(Lo): 43-74 mg/dL
L2(Hi): 288-389 mg/dL
03/17/17
Add Strip Lot
Place Meter above the Code
Key Reader and then press
the Right Arrow button to
start.
D-314: Do you want to
make Strip lot: 545603 the
'current' lot?
03/17/17
12:48 pm
12:48 pm
Make 'Current'
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator's Manual Version 6
Information about the expiration date and parameters for
control solutions is subsequently displayed.
1
Touch
to store the data for this lot number in
the meter without changes, or touch
the data for this lot number before storing it in the
meter.
If you accept the suggested values the following screen
appears. You can use this screen to select the lot number
you have just transmitted as the current lot number.
The current lot number is provided automatically for use
with subsequent tests.
2
Touch
to confirm that you want this lot
number to be the lot number currently in use, or
touch
to store the entries without making the
lot number the current lot number.
3
Continue entering additional lot numbers, or touch
to return to the Main Menu.
to modify