9.15 Beeper options

Main Menu
Patient Test
Control Test
Review Results
03/17/17
This setting can be used to set the volume of the beeper.
12:48 pm
Main Menu 2
Maintenance
Profi ciency
Strip Lots
Control Lots
03/17/17
1
2
3
4
08424705001 (01) 2017-04 USA • Accu-Chek Inform II Operator's Manual Version 6
12:48 pm
Beeper
Diagnostics
Linearity
Linearity Lots
Admin.
Touch
in the Main Menu to open the Main
Menu 2 screen.
Touch Beeper to set the volume.
Touch the button with the desired volume. When
you touch a button, the meter will beep at the
corresponding volume.
Touch
to save the setting and return to the
Main Menu 2 screen.
Beeper Setup
12:48 pm
Beeper Volume:
Low
Medium
High
03/17/17

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Inform II and is the answer not in the manual?

Questions and answers

askar
February 20, 2025

my code key reader does not work , how can i manually enter a new strip lot

1 comments:
Mr. Anderson
February 20, 2025

To manually enter a new strip lot on an Accu-Chek Inform II code key reader, follow these steps:

1. Open the Main Menu screen.
2. Select "Strip Lots" to access the related menu.
3. Touch "Add" to enter a new test strip lot from a new code key.
4. Insert the new code key into the code key reader.
5. The LED on the reader will flash green, indicating it is ready to transfer data.
6. The reader remains ready to send data for a few seconds after transmission, allowing multiple meters to receive the data without reinserting the code key.

This completes the manual entry of a new strip lot using the code key reader.