Power & energy meter
Hide thumbs Also See for Plus 2:

Advertisement

2.5. C
HOOSING MANUAL OR AUTOMATIC RANGING IN POWER MEASUREMENT
Touch the Range button
option.
Select the appropriate range manually, or choose the "Auto" mode
manage the range automatically.
2.6. O
FFSETTING CURRENT READING
You can activate the offset by touching the "Zero" button
the Plus 2.
When "offset" is activated, the button
its unit of measure.
To cancel the offset, press "offset" again.
2.7. Z
A
ERO
DJUSTMENTS
In the Plus 2, all adjustments, including ADC zeroing, are done by the software. It is
recommended to re-zero the Plus 2 each time the measuring head is changed.
a)
Press for more than 3 seconds the Zero button
on the screen of the Plus 2.
b) After 3 seconds, the confirmation pop-up message
(Figure 5) will be displayed.
Note:
While zeroing make sure that the instrument is not affected by
electrical noise and the sensor is undisturbed.
c) Press "Yes" and wait few seconds while the "Wait
for zeroing" and "Zeroing Completed" messages are
displayed (Figure 6).
LaserPoint s.r.l. - Plus 2 - Quick Start Guide - All Rights Reserved
on the screen of the Plus 2 to scroll the different ranging
:
shows a yellow led and the offset value with
:
to let the Plus 2
on the main screen of
Figure 5
Figure 6