Main electronic module
Levelflex FMP5x, Micropilot FMR5x
EA00041F/00/A2/15.14
71248975
Instruction is valid for the following spare part sets:
Order number
Device component
XPF0001 - ###...
Main electronic module for Levelflex FMP5x
XPF0020 - ###...
Main electronic module for Micropilot FMR5x
We recommend that the Installation Instructions be kept with the packaging at all times.
Confirmation whether the spare part is permitted to be used with the measuring device
The spare parts set and Installation Instructions are used to replace a faulty unit with a functioning unit of the same type. 
Use genuine parts from Endress+Hauser only.
Only original spare parts supplied by Endress+Hauser shall be used with the measuring device. Therefore, before use, check whether the spare
part set is compatible with the measuring device.
A spare parts overview label is located in the connection compartment cover of the measuring device. If there is no label or the spare part is not
listed the appropriate spare part can also be identified via W@M Device Viewer.
How carrying out such a check via the W@M Device Viewer is described below:
i
Choose the Endress+Hauser Device Viewer via web browser: www.endress.com/deviceviewer
Enter the serial number (Ser. No.) of the device into the W@M Device Viewer, then click on "Spare parts".
The list of the available spare parts for the device is displayed.
Check the order number on the packaging label of the spare part set.
Check whether the order number of the spare part set is specified in the list displayed by the device viewer:
= YES, the spare part can be used.
= NO, the spare part may not be used.
Ser. No.:12345
Ser.No.:1234567890
i
www.endress.com/deviceviewer
Device Viewer
Ser. No.: 12345
=
?
71248975

Advertisement

loading
Need help?

Need help?

Do you have a question about the Micropilot FMR5 series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Endress+Hauser Micropilot FMR5 series

  • Page 1 The spare parts set and Installation Instructions are used to replace a faulty unit with a functioning unit of the same type.  Use genuine parts from Endress+Hauser only. Only original spare parts supplied by Endress+Hauser shall be used with the measuring device. Therefore, before use, check whether the spare part set is compatible with the measuring device.
  • Page 2: Safety Instructions

    (for Ex. IECEx, ATEX, FM, CSA, TIIS, NEPSI) 1*, 2, 3 1 Trained customer technician 1*Trained customer technician, trained by Endress+Hauser (for repairs carried out on devices with type approval) 2 Service technician authorized by Endress+Hauser 3 Endress+Hauser (send measuring device back to manufacturer) Safety instructions •...
  • Page 3: Tool List

    Main electronic module Tool list 3 mm Replacing the main electronic module Note! De-energize the device before replacing the main electronic module (e.g. disconnect the terminal). The device can be damaged if the main electronic module is replaced while voltage is still applied. 3 mm 2.5 Nm Grafik Deckel...
  • Page 4 Main electronic module Endress + Hauser...
  • Page 5 Ersatzteilset und Einbauanleitung dienen dazu, eine defekte Einheit gegen eine funktionierende Einheit des gleichen Typs zu ersetzen.  Es dürfen nur Originalteile von Endress+Hauser verwendet werden. Grundsätzlich dürfen nur Ersatzteilsets verwendet werden, die von Endress+Hauser für das Messgerät vorgesehen sind. Vor der Verwendung ist deshalb zu überprüfen, ob das Ersatzteilset zum Messgerät passt.
  • Page 6 Zulassung (z.B. IECEx, ATEX, FM, CSA, TIIS, NEPSI) 1*, 2, 3 1 Ausgebildete Fachkraft des Kunden 1*Ausgebildete Fachkraft des Kunden, von Endress+Hauser geschult (für Reparaturen, die an Geräten mit Bauartzulassung vorgenommen werden) 2 Von Endress+Hauser autorisierter Servicetechniker 3 Endress+Hauser (Messgerät an Hersteller zurücksenden) Sicherheitshinweise •...
  • Page 7 Hauptelektronikmodul Werkzeugliste 3 mm Austausch Hauptelektronikmodul Hinweis! Vor einem Austausch des Hauptelektronikmoduls das Gerät spannungsfrei schalten (z.B. Abziehen der Anschlussklemmen). Das Messgerät kann beschädigt werden, wenn das Hauptelektronikmodul unter Spannung ausgetauscht wird. 3 mm 2.5 Nm Grafik Deckel Grafik Display1 Grafik Display2 Grafik Display3 0.5 Nm...
  • Page 8 Hauptelektronikmodul Endress + Hauser...