POOL EXACT EZ OVERVIEW ® YOUR NEW POOL EXACT EZ PHOTOMETER IS IDEAL FOR TESTING AND MAINTAINING YOUR POOL AND SPA WATER. SAMPLE CELL - 4 ML LIGHT SOURCE Built-in plastic 4mL cell Uses 525nm for collecting your water wavelength and...
FILL, DIP, READ TOTAL ALKALINITY POWER ON PHOTOMETER Press the button to power on the Pool eXact ® FILL CELL Rinse the cell 3 times with the water sample to FILL be tested and to the top to begin test.
FILL, DIP, READ TOTAL ALKALINITY REMOVE STRIP Remove one eXact Strip Micro AL Total Alkalinity ® and set in a dry, convenient place. Replace cap on bottle. DIP STRIP AND PRESS READ Press to initiate a 20 second countdown and...
Page 6
FILL, DIP, READ POWER ON PHOTOMETER Press the button to power on the Pool eXact ® FILL CELL Rinse the cell 3 times with the water sample to FILL be tested and to the top to begin test. Use the included cleaning brush to clean the cell before/ after each test to prevent reagent interferences and achieve the most accurate result.
Page 7
FILL, DIP, READ REMOVE STRIP Remove one eXact Strip Micro PH and set in a dry, ® convenient place. Replace cap on bottle. DIP STRIP AND PRESS READ Press to initiate a 20 second countdown and simultaneously the eXact Strip by submerging all ®...
FILL, DIP, READ FREE CHLORINE AND BROMINE POWER ON PHOTOMETER Press the button to power on the Pool eXact ® FILL CELL Rinse the cell 3 times with the water sample to FILL be tested and to the top to begin test.
Page 9
FILL, DIP, READ FREE CHLORINE AND BROMINE REMOVE STRIP Remove one eXact Strip Micro CL (DPD-1) Free ® Chlorine and set in a dry, convenient place. Replace cap on bottle DIP STRIP AND PRESS READ Press to initiate a 20 second countdown and...
Use the included cleaning brush to clean the cell before/ after each test to prevent reagent interferences and achieve the most accurate result. REMOVE STRIP Remove one eXact Strip Micro CL (DPD-3) Combined ® Chlorine and set in a dry, convenient place. Replace cap on bottle.
READ RESULTS READ result displayed as Combined Chlorine. This result is automatically stored in the CL3 menu. After testing is complete, rinse the sample cell immediately. eXact Strip Micro CL (DPD-1, DPD-3, DPD-4) Interferences ® Interfering Interfering Levels & Treatments...
FILL, DIP, READ PHOSPHATE POWER ON PHOTOMETER Press the button to power on the Pool eXact ® Clean CELL with 0.1N HCl, Distilled Vinegar (5%), or Muriatic Acid before fi lling the meter with the sample to be tested. FILL CELL...
Page 13
FILL, DIP, READ PHOSPHATE REMOVE STRIP Remove one eXact Strip Micro PO Phosphate and set ® in a dry, convenient place. Replace cap on bottle DIP STRIP AND PRESS READ Press to initiate a 20 second countdown and simultaneously the eXact Strip by submerging all ®...
FILL, DIP, READ CALCIUM HARDNESS POWER ON PHOTOMETER Press the button to power on the Pool eXact ® FILL CELL Rinse the cell 3 times with the water sample to FILL be tested and to the top to begin test.
FILL, DIP, READ CALCIUM HARDNESS REMOVE STRIP Remove one eXact Strip Micro CA Calcium ® Hardness and set in a dry, convenient place. Replace cap on bottle. DIP STRIP AND PRESS READ Press to initiate a 20 second countdown and...
FILL, DIP, READ CYANURIC ACID POWER ON PHOTOMETER Press the button to power on the Pool eXact ® FILL CELL Rinse the cell 3 times with the water sample to FILL be tested and to the top to begin test.
Page 17
CYANURIC ACID SHAKE BOTTLE & ADD DROPS Tilt the meter to discard about 0.2mL water to leave room for liquid reagent. Shake the bottle of eXact ® Reagent CY to mix the suspension in the bottle. Add 5 drops of the eXact Reagent CY to the sample cell.
FILL, DIP, READ COPPER POWER ON PHOTOMETER Press the button to power on the Pool eXact ® FILL CELL Rinse the cell 3 times with the water sample to FILL be tested and to the top to begin test. Use the included cleaning brush to clean the cell before/ after each test to prevent reagent interferences and achieve the most accurate result.
Page 19
FILL, DIP, READ COPPER REMOVE STRIP Remove one eXact Strip Micro CU Copper and set in a ® dry, convenient place. Replace cap on bottle DIP STRIP AND PRESS READ Press to initiate a 20 second countdown and simultaneously the eXact Strip by submerging all ®...
Page 20
CHLORIDE (Salt) PREPARE SAMPLE Prepare sample for testing using Mini Dilution Kit II #487202. 1. Rinse 50mL graduated conical tube with distilled or deionized (salt-free) water. 2. Rinse the 3.0mL syringe with saltwater sample to be tested. Finally, fi ll the 3.0mL syringe to the 2.0mL line very precisely (plunger ring should line up at the 2.0mL line and little or no air bubble should be present).
FILL, DIP, READ CHLORIDE (Salt) POWER ON PHOTOMETER Press the button to power on the Pool eXact ® FILL CELL Rinse the cell 3 times with the water sample to FILL be tested and to the top to begin test.
Page 22
‘1’. Remove and discard the strip. READ RESULTS READ result displayed as Chloride. Due to the limit of three digits on the Pool eXact EZ photometer, you ® should add a “0” to the end of the result on the LCD (ex.
FOR BEST ACCURACY The Pool eXact EZ photometer has a 5 minute auto-shutoff timer. ® Each test menu can store 20 results. To retrieve the stored results, go to the desired test using the MENU key. When the desired test is displayed, press and hold down the MENU key.
The Pool eXact ® EZ has been factory calibrated and will stay valid because of its exceptional quality. We are so confi dent in the Pool eXact ® we offer an industry leading 2-year warranty. We built the Pool eXact EZ to be easy, accurate and ®...
Registration is available over the phone (+1-803-329-9712 Ext. 0) or online at sensafe.com/micro/ warranty/ (Personal data is kept confi dential). The Pool eXact EZ photometer ® is warranted to be free from defects in materials and workmanship for a period of two (2) years from the date of purchase by the customer.
38 408 G4/G5, ISO 7393/2) accepted for testing requirements for Free Chlorine and Total Chlorine. The compliance requirement is a photometer wavelength to measure between 490 and 530nm. The Pool eXact EZ smart ® photometer uses a 525nm wavelength and 11 mm path-length.
Performance verifi ed with various salt systems and water samples with optimal water temperature at 10-40˚C / 50-104˚F. Optimal water temperature for Phosphate test is 15-40˚C / 59-104˚F. For detailed instructions on using this product, contact us (see below) or visit sensafe.com/photometers/pool-exact-ez/. Requires the use of 2 strips if reading is above 6 ppm.
Need help?